Chapter 6 151
Advanced Operations
Status Reporting
Calibration Status Register Group
The Calibration Status Register Group contains information about the
Test Set’s hardware. This status group is accessed using the STATus
commands. The Calibration Status Register Group uses 16-bit registers
and includes a Condition Register, Transition Filters, an Event
Register, an Enable Register, and a Summary Message. Refer to the
“Status Reporting Structure Overview” on page 121 for a discussion of
status register operation. Figure 6-11 shows the structure and STATus
commands for the Calibration Status Register Group.
Figure 6-11 Calibration Status Register Group
0
:STATus: CALibration: ENABle
?
<integer>
:STATus: CALibration: EVENt ?
:STATus: CALibration
:PTRansition
:NTRansition
?
<integer>
?
<integer>
:STATus: CALibration: CONDition ?
Summary Message Bit
Condition
Register
Transition
Filter Registers
Event
Register
Enable
Register
1
2
3
4
5
6
7
8
9
10
11
12
13
14
L
o
g
i
c
a
l
O
R
15
ch4drw11.drw
(to bit 8 of Questionable
Data/Signal Register Group
Condition Register