162 Chapter 6
O:\Manuals\E6380A_Progguid\Book\Advoper.fm
Advanced Operations
Status Reporting
Example 6-86 Example
OUTPUT 714;"STAT:HARD1:PTR 256"
Reading the Event Register
Example 6-87 Syntax
STATus:HARDware1:EVENt?
Example 6-88 Example
OUTPUT 714;"STAT:HARD1:EVEN?"
ENTER 714;Register_value
Clearing the Event Register
The EVENT register is cleared whenever it is queried or whenever the
Common Command *CLS is sent to the Test Set.
Reading the Enable Register
Example 6-89 Syntax
STATus:HARDware1:ENABle?
Example 6-90 Example
OUTPUT 714;"STAT:HARD1:ENAB?"
ENTER 714;Register_value
Writing the Enable Register
Example 6-91 Syntax
STATus:HARDware1:ENABle <integer>
Example 6-92 Example
OUTPUT 714;"STAT:HARD1:ENAB 256"
Clearing the Enable Register
The ENABLE register is cleared by writing to it with an integer value