Agilent Technologies 8935 Series Instrukcja Użytkownika Strona 141

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 329
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 140
Chapter 6 141
Advanced Operations
Status Reporting
Example 6-34 Example
OUTPUT 714;"*ESE?"
ENTER 714;Register_value
The *ESE? query allows the programmer to determine the current
contents (bit pattern) of the Standard Event Status Enable Register.
The Test Set responds to the *ESE? query by placing the
binary-weighted decimal value of the Standard Event Status Enable
Register bit pattern into the Output Queue. The decimal value of the bit
pattern will be a positive integer in the range of 0 to 255. The response
data is obtained by reading the Output Queue into a numeric variable,
integer or real.
Example 6-35 Example BASIC program
10 INTEGER Std_evn_enab_rg
20 OUTPUT 714;"ESE?"
30 ENTER 714;Std_evn_enab_rg
40 PRINT Std_evn_enab_rg
50 END
Example 6-36 Example response
36
Writing the Enable Register
Example 6-37 Syntax
*ESE <integer>
Przeglądanie stron 140
1 2 ... 136 137 138 139 140 141 142 143 144 145 146 ... 328 329

Komentarze do niniejszej Instrukcji

Brak uwag