Agilent Technologies N9344C Instrukcja Obsługi Strona 193

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 224
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 192
Command Reference 5
Programmer’s Guide 185
ASK Measurement Length
[:SENSe]:DMA:MEASurement:LENGth <value>
[:SENSe]:DMA:MEASurement:LENGth?
Use this command to specify measurement length of ASK demodulation
analysis.
Range: 20 to 1100
Example: :DMA:MEASurement:LENGth 200
Key access: Meas > Meas Length
ASK Trigger Type
:TRIGger:DMA[:SEQuence]:SOURce IMMediate|EXTer-
nal|RFTRigger
:TRIGger:DMA[:SEQuence]:SOURce?
Use this command to specify the trigger type used in current ASK
demodulation analysis.
Range: IMMediate, EXTernal, RFTRigger
Example: :TRIGger:DMA:SOURce EXTernal
Key access: Meas > Trigger
ASK RF Trigger Level
:TRIGger:DMA[:SEQuence]:LEVel:ABSolute <value>
:TRIGger:DMA[:SEQuence]:LEVel:ABSolute?
Use this command to turn on/off scale coupling state of X axis.
Range: 40 to 10 dBm
Example: :TRIGger:DMA:LEVel:ABSolute 5dBm
Key access: Meas > Trigger > RF Trigger > Video Level
Przeglądanie stron 192
1 2 ... 188 189 190 191 192 193 194 195 196 197 198 ... 223 224

Komentarze do niniejszej Instrukcji

Brak uwag