AM X Scale/DIV
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:PDIVi-
sion < value >
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:PDIVi-
sion?
Use this command to set the Scale per division of X axis. The unit of
this parameter is second. This command is only available when the scale
coupling state is set to off.
Range: 1e–9 to 1
Example: :DISPlay:AMA:VIEW:WINDow:TRACe:X:PDIVision 2e4
Key access: Shift > Disp >X Scale > Scale/DIV
AM X Reference Value
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:RVALue
< value >
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:RVALue?
Use this command to set the reference value of X axis.
Range: –5 to 5 s
Example: :DISPlay:AMA:VIEW:WINDow:TRACe:X:RVALue 1
Key access: Shift > Disp >X Scale > Ref Value
AM X Reference Position
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:RPOSi-
tion
LEFT|CENTer|RIGHt
:DISPlay:AMA:VIEW:WINDow:TRACe:X[:SCALe]:RPOSi-
tion?
Use this command to set the reference position of X axis.
Range: LEFT, CENTer, RIGHt
Example: :DISPlay:AMA:VIEW:WINDow:TRACe:X:RPOSition CENT
Key access: Shift > Disp >X Scale > Ref Position