120 Programmer’s Guide
5 Command Reference
Query SEM Marker Y Value
[:SENSe]:SEMask:MARKer[1]|2|3|4|5|6:Y?
Use this command to query the marker readout of Y axis in SEM
measurement.
Example: :SEMask:MARKer1:Y?
SEM Marker All Off
[:SENSe]:SEMask:MARKer:AOFF
Use this command to turn off all the marker in SEM measurement.
Example: :SEMask:MARKer:AOFF
Key access: Marker > Off
SEM Offset Resolution BW
[:SENSe]:SEMask:OFFSet:LIST:BAND-
width|BWIDth[:RESolution] <real_number>,
<real_number>, <real_number>,
<real_number>,<real_number>
[:SENSe]:SEMask:OFFSet:LIST:BAND-
width|BWIDth[:RESolution]?
Use this command to specify the resolution bandwidth of reference
channels in SEM measurement.
[:SENSe]:SEMask:OFFSet:LIST:BANDwidth|BWIDth[:RESolution]:
AUTO OFF|ON|0|1,...
[:SENSe]:SEMask:OFFSet:LIST:BANDwidth|BWIDth[:RESolution]:
AUTO?
Use this command to turn on/off automatically setting of resolution
bandwidth of reference channel in SEM measurement.
Example: SENSe:SEMask:OFFSet:LIST:BANDwidth 1E6, 1E6,
1E6, 1E6,1E6
Key access: Meas > Offset/Limit > Res BW