AC Current
Measurement Method: DC coupled to the fuse and current shunt, AC coupled true
rms measurement (measures the AC component only)
Shunt Resistance: 1 Ω for 10 mA, 100 mA
0.01 Ω for 1 A, 10 A
Input Protection: Externally accessible at rear panel 0.4 A, 500 V fuse
for I terminal
Internal 11 A, 1000 V fuse for 10 A terminal
Frequency
Measurement Method: Reciprocal counting technique. AC coupled input using AC
voltage function.
Signal Level: 10% of range to full scale input on all ranges except
where noted.
Auto or manual range selection
Gate Time: 0.1 second or 1 period of the input signal.
Input Protection: 750 V rms on all ranges (HI terminal)
Math Functions Null, dBm, dB, Min/Max/Avg, Hold, Limit Test
Data Log Info, List, Histrogram
Triggering and Memory
Samples per Trigger:
Trigger Delay:
1 to 5,000 (typical), 1 to 50,000 (optional)
0 to 3600 sec (100 us step size)
Trigger Out
3.3 V Logic output
Polarity:
Pulse width:
Negative pulse
Approximately 3 μs
Non-volatile Memory 50,000 readings
Sample Timer Range: Up to 3600 sec in 100 μs steps
Remote Interface USB 2.0 Standard, GPIB IEEE-488 (Optional)
Programming Language SCPI-1994.0, IEEE-488.2
Table 7 .Supplemental measurement specifications (continued)