5
AC Specifications
Table 2. AC Accuracy ± (% of reading + % of range)
Function Range
1
Frequency
1 Year
23 °C ± 5 °C
Temperature Coefficient
0 °C - 18 °C
28 °C - 55 °C
True RMS AC Voltage
2
100.000 mV
20 Hz – 45 Hz 1.0 + 0.1 0.02 + 0.02
45 Hz – 10 kHz 0.2 + 0.1 0.02 + 0.02
10 kHz – 30 kHz 1.5 + 0.3 0.05 + 0.02
30 kHz – 100 kHz
3
3.0 + 0.3 0.10 + 0.02
1.00000 V to 750.00 V
20 Hz – 45 Hz 1.0 + 0.1
4
0.02 + 0.02
45 Hz – 10 kHz 0.2 + 0.1 0.02 + 0.02
10 kHz – 30 kHz 1.5 + 0.3 0.05 + 0.02
30 kHz – 100 kHz
3
3.0 + 0.3
5
0.10 + 0.02
True RMS AC Current
2
10.0000 mA to 10.0000 A
20 Hz – 45 Hz 1.5 + 0.1 0.02 + 0.02
45 Hz – 1 kHz 0.5 + 0.1 0.02 + 0.02
1 kHz – 10 kHz
6
2.0 + 0.2 0.02 + 0.02
Notes on Specifications:
1. 20% over range on all ranges except ACV 750 V and ACI 10 A
2. Specifications are for sine-wave inputs more than the 5 % of range except 750 V range. Input signal must be more than 50 Vrms for 750 V range.
Maximum crest factor of 3 at full scale. Input impedance is 1 MΩ in parallel with capacitance less than 120 pF, AC couple with up to 400 DCV
3. Additional error to be added as frequency > 30 kHz and signal input < 10% of range. 30 kHz to 100 kHz: 0.003% of full scale per kHz
4. For input < 200 V rms
5. For input < 300 V rms
6. Frequencies > 5 kHz are typical for all ranges
Table 3. Frequency Accuracy ± (% of reading + 3 counts)
Function Range
1
Frequency
1 Year
23 °C ± 5 °C
Temperature Coefficient
0 °C - 18 °C
28 °C - 55 °C
Frequency
100.000 mV to 750.00 V
1
20 Hz – 300 kHz
2
0.02 + 3 0.005
10.0000 mA to 10.0000 A 20 Hz – 10 kHz
3
0.02 + 3 0.005
Notes on Specifications:
1. The frequency can be measured up to 1 MHz as 0.5 V signal to 100 mV / 1 V ranges.
2. 10% of range to full scale input on all ranges except where noted. 100 mV range specifications are for full scale or greater inputs.
For inputs from 10 mV to 100 mV, multiply total % of reading error by 10.
3. 10% of range to full scale input on all ranges except where noted. 10 mA range specifications are for full scale or greater inputs.
For inputs from 1 mA to 10 mA, multiply total % of reading error by 10.
Table 4. Frequency resolution
Function Range Frequency Resolution
Frequency 100.000 mV to 750.00 V
1
119.999 Hz 0.001 Hz
1.19999 kHz 0.00001 kHz
11.9999 kHz 0.0001 kHz
119.999 kHz 0.001 kHz
1.19999 MHz 0.00001 MHz
1. The frequency can be measured up to 1 MHz as 0.5 V signal to 100 mV / 1 V ranges.
Specifi cations are for 90 minutes warm- up time, slow
mode, and calibration temperature within 18 °C - 28 °C