Supplemental Measurement Specifications
Table 7 .Supplemental measurement specifications
DC Voltage
Measuring Method: Sigma Delta A-to-D converter
Input Resistance: > 10 GΩ ± 2% range (Selectable 100 mV, 1 V ranges)
10 MΩ ± 2% range (typical)
Input Protection: 1000 V on all ranges (HI terminal)
Resistance
Measurement Method: 2-wire or 4-wire Ohms
Open circuit voltage: Limited to < 2.8 V
Input Protection: 1000 V on all ranges (HI terminal)
DC Current
Shunt Resistance: 1 Ω for 10 mA, 100 mA
90 Ω for 100 μA, 1mA
0.01 Ω for 1 A, 10 A
Input Protection: Externally accessible at Rear Panel 0.4 A, 500 V fuse
for I terminal
Internal 11 A, 1000 V fuse for 10 A terminal
Continuity / Diode Test
Measurement Method: Uses 0.5 mA ± 0.2% constant current source
Response Time: Continuity : 165 samples/second with audible tone
Diode : 190 samples/second with audible tone
Continuity Threshold: 10 Ω fixed
Input Protection: 1000 V (HI terminal)
Temperature
Measurement Method: 2-wire Ohms measurement of 5 kΩ thermistor sensor
(YSI 4407) with computer conversion
Auto-ranging measurement, no manual range selection
Input Protection: 1000 V (HI terminal)
Measurement Noise
Rejection
CMR (Common Mode Rejection)
For 1 kΩ unbalance LO lead
DC 140 dB
AC 70 dB
NMR (Normal Mode Rejection)
For 60 Hz (50 Hz) ± 0.1%
Slow mode 5½ digits 90 dB
Medium mode 4½ digits 55 dB
Fast mode 4½ digits 0 dB
AC Voltage
Measurement Method: AC coupled true-rms - measures the AC component with up
to 400 VDC bias any range
Crest Factor: Maximum 3:1 at full scale
Input Impedance: 1MΩ ± 2% in parallel with < 100 pF of all ranges
Input Protection: 750 V rms on all ranges (HI terminal)