8
Measurement Supplemental Characteristics
Temperature coefficient (0 °C to 18 °C and 28 °C to 50 °C):
±(0.1 x Accuracy specification) /°C.
Over range :
102 % of measurement range for 1.5 A and 3 A ranges. 106 % of measure
ment range other than 1.5 A and 3 A ranges.
Voltage measurement range change overshoot:
< 250 mV. 100 kΩ load, 20 MHz bandwidth
Current measurement range change overshoot:
<250 mV/R load , 20 MHz bandwidth
Derating accuracy for measurement speed less than 1 PLC :
Add % of range using the following table for measurement with PLC <1.
Derating accuracy with PLC setting< 1 PLC
Triggering and Timing Specifications
Triggering
Digital I/O Trigger in to trigger out
1
: ≤ 5 μs
Digital I/O Trigger in to source change
1
: ≤ 5 μs
LXI Trigger in to source change
1
: Minimum 100 μs, Typical 200 μs, Maximum
Unknown
LXI Trigger in to measurement
1
: Minimum 100 μs, Typical 200 μs, Maximum
Unknown
Internal event to external LXI Trigger output
1
: Minimum 100 μs, Typical 200 μs,
Maximum Unknown
LAN[0-7] Event Receiving Latency
1
: Unknown
LAN[0-7] Event Sending Latency
1
: Unknown
Minimum trigger interval: 10 μs
1.Supplemental Characteristics
Timer
Trigger timing resolution: 1 μs to 100 ms
Timer accuracy: ±50 ppm
Arm/trigger delay: 0 to 100000 s
Arm/trigger interval:
B2901A/B2902A: 20 μs to 100000 s
B2911A/B2912A: 10 μs to 100000 s
Arm/trigger event: 1 to 100000
Time stamp:
TIMER value automatically saved when each measurement is triggered.
Voltage Range Current Range
0.2 V 2 V to
200 V
10 nA 100 nA 1 μA to
100 mA
1 A to
3 A
0.1 PLC 0.01 % 0.01 % 0.1 % 0.01 % 0.01 % 0.01 %
0.01 PLC 0.05 % 0.02 % 1 % 0.1 % 0.05 % 0.02 %
0.001
PLC
0.5 % 0.2 % 5 % 1 % 0.5 % 0.2 %