5
Source Specifications
DC Voltage source specifications
Range
Programming
Resolution
Accuracy
(% reading + offset)
Noise
(peak to peak)
0.1 Hz to 10 Hz
1
Maximum Voltage
(over range)
±200 mV 100 nV ±(0.015 % + 225 μV) ≤ 5 μV ±210 mV
±2 V 1 μV ±(0.02 % + 350 μV) ≤ 15 μV ±2.1 V
±20 V 10 μV ±(0.015 % + 5 mV) ≤ 150 μV ±21 V
±200 V 100 μV ±(0.015 % + 50 mV) ≤ 1.5 mV ±210 V
1. Supplemental characteristics
Range
Programming
Resolution
Accuracy
(% reading + offset)
Noise
(peak to peak)
0.1 Hz to 10 Hz
1
Maximum Current
(over range)
±10 nA 10 fA ±(0.10 % + 50 pA) ≤ 1 pA ±10.5 nA
±100 nA 100 fA ±(0.06 % + 100 pA) ≤ 2 pA ±105 nA
±1 μA 1 pA ±(0.025 % + 500 pA) ≤ 20 pA ±1.05 μA
±10 μA 10 pA ±(0.025 % + 1.5 nA) ≤ 60 pA ±10.5 μA
±100 μA 100 pA ±(0.02 % + 25 nA) ≤ 1 nA ±105 μA
±1 mA 1 nA ±(0.02 % + 200 nA) ≤ 6 nA ±1.05 mA
±10 mA 10 nA ±(0.02 % + 2.5 μA) ≤ 100 nA ±10.5 mA
±100 mA 100 nA ±(0.02 % + 20 μA) ≤ 600 nA ±105 mA
±1 A 1 μA ±(0.03 % + 1.5 mA) ≤ 20 μA ±1.05 A
±1.5 A 1 μA ±(0.05 % + 3.5 mA) ≤ 20 μA ±1.515 A
±3 A 10 μA ±(0.4 % + 7 mA) ≤ 60 μA ±3.03 A
±10 A
2
10 μA ±(0.4 % + 25 mA)
3
±10.5 A
DC Current source specifications
1. Supplemental characteristics
2. 10 nA range is available only for pulse mode, not available for DC mode.
3. Measurement speed: 0.01 PLC