Agilent Technologies B2902 Instrukcja Użytkownika Strona 26

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 35
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 25
10
Slew rate/small signal bandwidth
Voltage Range Load
Resistance
Current range 200 mV 2 V 20 V 200 V
1 mA 94 µA/µs, 21 kHz 160 µA/µs, 21 kHz 160 µA/µs, 21 kHz 150 µA/µs, 21 kHz 100 Ω
10 mA 94 µA/µs, 21 kHz 670 µA/µs, 21 kHz 900 µA/µs, 21 kHz 900 µA/µs, 21 kHz 100 Ω
100 mA 8 mA/µs, 10 kHz 8 mA/µs, 10 kHz 8 mA/µs, 10 kHz 12 mA/µs, 21 kHz 1 Ω
1 A 78 mA/µs, 12 kHz 94 mA/µs, 12 kHz 92 mA/µs, 12 kHz 0.1 Ω
1.5 A 125 mA/µs, 12 kHz 135 mA/µs, 12 kHz 140 mA/µs, 12 kHz 0.1 Ω
3 A 250 mA/µs, 13 kHz 270 mA/µs, 13 kHz 260 mA/µs, 13 kHz 0.041 Ω
Note: Slew rate and small signal bandwidth can be down to -20% in maximum.
I source overshoot:
< ±0.1 % (<± 0.3 % for 3 A range). Step is 10 % to 90 % range, resistive load.
Current source range change overshoot:
≤ 250 mV/R load , 20 MHz bandwidth
Line regulation/load regulation: Included in current source specifications.
High Capacitance Mode
The high capacitance mode supports sources and measurements when the load
capacitance is greater than 0.01 μF. In high capacitance mode the maximum
allowed load capacitance value is 50 μF.
Voltage output settling time:
Time required to reach within 0.1 % of final value with 4.7 μF capacitive load
on a fixed range at specified current range and limit value.
200 mV, 2 V Ranges: 600 μs, at 1 A limit
20 V Range: 1.5 ms, at 1 A limit
200 V Range: 20 ms, at 100 mA limit
Current measurement settling time:
Time required to reach 0.1 % of the final value after the voltage source stabi
lizes in a fixed range. Vout is 5 V unless noted.
1 μA Range: 230 ms
10 μA, 100 μA Ranges: 23 ms
1 mA, 10 mA Ranges: 0.23 ms
100 mA to 3 A Ranges: 100 μs
Mode change delay:
Delay into High Cap mode:
1 μA Range: 230 ms
10 μA, 100 μA Ranges: 23 ms
1 mA to 3 A Ranges: 1 ms
Delay out of High Cap mode:
All ranges: 10 ms
Noise 10 Hz to 20 MHz (20 V range): 4.5 mVrms
Voltage source range change overshoot (20 V range or below):
< 250 mV, 20 MHz bandwidth
High Capacitance mode working conditions:
V/I mode: Voltage source mode only
Range: Current measurement range is limited to fixed range only. 10 nA and
100 nA ranges are not available.
Current limit: ≥ 1 µA
Przeglądanie stron 25