12
Settling time
Voltage source
1
: 80 ms (2-wire), 140 ms (4-wire)
Current source
2
: 11 ms (2-wire), 150 ms (4-wire)
1. Time required to reach within 0.1 % of final value at open load condition. Step is 10 % to 90 %
range. At 20 V range, 100 mA limit/100 mA range
2. Time required to reach within 0.1 % of final value at short condition. Step is 10 % to 90 % range.
At 100 mA range, 20 V limit / 20V range
Load transient recovery time
Time to recover to within the settling band following a load change.
50 mA current change
1
Settling band ± 20 mV
Time 50 ms (2-wire), 85 ms (4-wire)
1. At a load change to change the flowing current from +10 mA to +60 mA, 10 V voltage force, 20 V
range, 100 mA limit.
Supported ranges:
Voltage: 200 mV to 200 V ranges (42 V maximum)
Current: 10 µA to 100 mA ranges (105 mA maximum)
Maximum capacitive load: 50 µF (for 4-wire)
Dimensions: 41 mm H x 58.2 mm W x 141.5 mm D (When the filter is inserted to
the output connector of B2961A/B2962A, the depth is 126.5 mm.)
Weight: 0.3 kg
Note:
* The 10 A pulse range and programmable output resistance capability are not
supported by external filters.
* For the current output, only DC is supported when using with external filters.
*The current measurement data monitored by the built-in voltage/current moni
tor can be influenced by the charge and discharge current of the capacitance
inside of the filters.
1.E-10
1.E-09
1.E-08
1.E-07
1.E-06
1.E-05
1.E-04
1.E+00 1.E+01 1.E+02 1.E+03 1.E+04 1.E+05 1.E+06 1.E+07 1.E+08
Noise Density [Vrms/√Hz]
Frequency [Hz]
Noise Density
20V range
2V range