Agilent B2900 SCPI Command Reference, Edition 2 5- 13
Error Messages
B2900 Specific Error
Error 134 Self-test failed; DAC/ADC, item; channel n
Failed the DAC/ADC test specified by item and channel n.
Error 135 Self-test failed; Loop control, item; channel n
Failed the loop control test specified by item and channel n.
Error 136 Self-test failed; I sense, item; channel n
Failed the current sense test specified by item and channel n.
Error 137 Self-test failed; V sense, item; channel n
Failed the voltage sense test specified by item and channel n.
Error 138 Self-test failed; F-COM comparison, item; channel n
Failed the F-COM comparison test specified by item and channel n.
Error 139 Self-test failed; V switch, item; channel n
Failed the voltage switch test specified by item and channel n.
Error 140 Self-test failed; Temperature sensor, item; channel n
Failed the temperature sensor test specified by item and channel n.
Error 202 Instrument locked by another I/O session
The requested operation is not allowed because the instrument is locked by another
I/O session. The instrument must be unlocked.
Error 210 Operation is not completed
Operation is still in progress. Wait for operation complete.
Error 211 Cannot switch low sense terminal with output on
Output relay must be off to switch low sense terminal.
Error 212 Output relay must be on
Output relay must be on to perform the requested operation.
Error 213 Output relay must be off
Output relay must be off to perform the requested operation.