Agilent Technologies B2902 Instrukcja Obsługi Strona 253

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 326
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 252
Agilent B2900 SCPI Command Reference, Edition 2 4- 131
Subsystem Commands
:SYSTem:COMMunicate:LAN:<HNAMe|HOSTname>
Parameter address IP address of the default gateway. It must be in the A.B.C.D
format with 15 characters maximum. A, B, C, and D must be a
number from 0 to 225. Parameter data type is SPD.
Query response address <newline>
address returns the IP address of the default gateway. If the CURRent parameter is
set, address returns the present setting. If the STATic parameter is set, address
returns the reserved value for the next startup. Response data type is SRD.
Example :SYST:COMM:LAN:GATE “192.168.100.210”
:SYST:COMM:LAN:GATE?
:SYSTem:COMMunicate:LAN:<HNAMe|HOSTname>
Sets the host name of the instrument. The setting is enabled by the
:SYSTem:COMMunicate:LAN:UPDate command. This command setting is not
changed by power off or the *RST command.
Syntax :SYSTem:COMMunicate:LAN:<HNAMe|HOSTname> hostname
:SYSTem:COMMunicate:LAN:<HNAMe|HOSTname>? [CURRent|STATic]
Parameter hostname Host name. Up to 15 characters. Parameter data type is SPD.
Query response hostname <newline>
hostname returns the host name of the instrument. If the CURRent parameter is set,
hostname returns the present setting. If the STATic parameter is set, hostname
returns the reserved value for the next startup. Response data type is SRD.
Example :SYST:COMM:LAN:HNAM “A-B2911A-00001”
:SYST:COMM:LAN:HOST?
:SYSTem:COMMunicate:LAN:MAC?
Returns the MAC address of the instrument.
Syntax :SYSTem:COMMunicate:LAN:MAC?
Query response mac_address <newline>
Przeglądanie stron 252
1 2 ... 248 249 250 251 252 253 254 255 256 257 258 ... 325 326

Komentarze do niniejszej Instrukcji

Brak uwag