Agilent Technologies B2902 Instrukcja Obsługi Strona 159

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 326
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 158
Agilent B2900 SCPI Command Reference, Edition 2 4- 37
Subsystem Commands
:HCOPy:SDUMp:DATA?
HCOPy Subsystem
:HCOPy:SDUMp:DATA?
Returns the data of the front panel screen image. The format of the image data is set
by the
:HCOPy:SDUMp:FORMat command.
Syntax :HCOPy:SDUMp:DATA?
Query response The response is a definite length arbitrary binary block.
Example :HCOP:SDUM:DATA?
:HCOPy:SDUMp:FORMat
Sets the format of the image data. The front panel screen image will be created in
the format set by this command. The image data will be returned by the
:HCOPy:SDUMp:DATA? command.
Syntax :HCOPy:SDUMp:FORMat format
HCOPy:SDUMp:FORMat?
Parameter format Format of image data. JPG (default)|BMP|PNG|WMF.
Parameter data type is CPD.
Query response format <newline>
format returns JPG, BMP, PNG, or WMF. Response data type is CRD.
Example :HCOP:SDUM:FORM BMP
:HCOP:SDUM:FORM?
Przeglądanie stron 158
1 2 ... 154 155 156 157 158 159 160 161 162 163 164 ... 325 326

Komentarze do niniejszej Instrukcji

Brak uwag