4- 110 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
[:SOURce]:SWEep:SPACing
mode returns the present setting of the output ranging mode, BEST, FIX, or AUTO.
Response data type is CRD.
Example :SWE:RANG BEST
:SOUR2:SWE:RANG?
[:SOURce]:SWEep:SPACing
Selects the scale of the sweep output for the specified channel.
Syntax [:SOURce[c]]:SWEep:SPACing mode
[:SOURce[c]]:SWEep:SPACing?
Parameter mode Sweep scale. LOGarithmic|LINear (default). Parameter data
type is CPD.
mode=LIN selects the linear scale sweep output.
mode=LOG selects the logarithmic scale sweep output. For the log sweep, the
sweep step value is ignored.
Query response mode <newline>
mode returns the present setting of the scale, LIN or LOG. Response data type is
CRD.
Example :SWE:SPAC LOG
:SOUR2:SWE:SPAC?
[:SOURce]:SWEep:STAir
Sets the sweep mode for the specified channel.
Syntax [:SOURce[c]]:SWEep:STAir mode
[:SOURce[c]]:SWEep:STAir?
Parameter mode Sweep mode. SINGle (default)|DOUBle. Parameter data type is
CPD.
mode=SINGle sets the sweep mode to single sweep.