Agilent Technologies B1500A Podręcznik Użytkownika Strona 88

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 204
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 87
3- 10 Keysight B1500A User’s Guide, Edition 14
Installation
Inspection and Installation
To Change GPIB Address
When this instrument is shipped from the factory, the GPIB address is set to 17. To
change the GPIB address, perform the following procedure.
1. If the EasyEXPERT software is running, terminate it as shown below:
a. Select the menu function File > Exit on the EasyEXPERT main screen.
b. Click [x] at the upper right corner of the Start EasyEXPERT button.
2. Select All Programs > Keysight IO Libraries Suite > Keysight Connection
Expert from the Start menu. Keysight Connection Expert window appears.
3. At the Instrument I/O on this PC area, select GPIB0, and click the Change
Properties... button. Keysight 82350 PCI GPIB Interface - GPIB0 window
appears.
4. Change the GPIB Address value.
5. Remove the check from the Auto-discover instruments connected to this
interface box.
6. Click the OK button on Keysight 82350 PCI GPIB Interface - GPIB0 window.
7. If the Reboot Required dialog box is displayed, click the Reboot Now button,
and reboot the instrument.
Przeglądanie stron 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 203 204

Komentarze do niniejszej Instrukcji

Brak uwag