Keysight B1500A User’s Guide, Ed ition 14 1- 15
Getting Started
To Use Classic Test Mode
See “To Perform Measurement” on page 1-16 to start measurement.
Table 1-4 I/V Sweep Channel Setup
Table 1-5 I/V Sweep Measurement Setup
Table 1-6 I/V Sweep Display Setup
Unit V Name I Name
Mode
a
a. Channel output mode. Set I or IPULSE for the current output and volt-
age measurement. Set V, VPULSE, or COMMON for the voltage output
and current measurement.
Function
SMU1 V1 I1 V VAR1
SMU2 V2 I2 V CONST
VAR1 Timing Constants
Unit SMU1 Hold 0 s Unit SMU2
Name V1 Delay 0 s V Name V2
Direction Single I Name I2
Linear/Log LINEAR Mode V
Start 0 V Source 0 V
Stop 1 V Compliance 100 mA
Step 10 mV
No of Step 101
Compliance 100 mA
Pwr Comp OFF Sweep CONTINUE AT ANY status
Name Sharing Scale Min Max
X V1 (None) Linear 0 V 1 V
Y1 I1 (None) Linear -1 pA 1 pA