2- 26 Keysight B1500A User’s Guide, Edition 14
Introduction
Measurement Modules
MCSMU - Medium Current SMU
This section describes typical specification of the medium current source/monitor
unit (MCSMU). Maximum four modules can be installed in one mainframe.
Maximum current: ± 1 A (pulse), ± 0.1 A (DC)
Maximum voltage: ± 30 V
Maximum output power: 30 W (pulse), 3 W (DC)
Minimum range: 0.2 V, 10 A
Minimum pulse width: 10 s
Output/measurement value and resolution: see Table 2-9 and Table 2-10.
Figure 2-5 MCSMU Output and Measurement Ranges
30
30
1
0.1
0.1
1
Voltage (V)
Current (A)
-
-
-
Pulse mode only
Pulse mode only