Mainframe Specifications (continued)
Standard (Direct Trigger) Option 001 (Divided Trigger)
Trigger Modes
Internal Trigger
1
Freerun
External Direct Trigger
2
Limited Bandwidth
3
DC to 100 MHz
Full Bandwidth DC to 2.75 GHz
External Divided Trigger 2 to 12 GHz (1 to 15 GHz)
Jitter
Characteristic <1.0 ps RMS + 5*10E-5 of delay setting
4
1.2 ps RMS for time delays less than 100 ns
Maximum 1.5 ps RMS + 5*10E-5 of delay setting
4
1.7 ps RMS for time delays less than 100 ns
Trigger Sensitivity 200 m Vpp (sinusoidal input or 200 m Vpp sinusoidal input from 2 to 12 GHz
200 ps minimum pulse width)
Trigger Configuration
Trigger Level Adjustment –1 V to + 1 V AC coupled
Edge Select Positive or negative
Hysteresis
5
Normal or high sensitivity
Trigger Gating
Gating Input Levels Disable: 0 to 0.6 V
(TTL compatible) Enable: 3.5 to 5 V
Pulse width >500 ns, period >1 µs
Gating Delay Disable: 627 ns + trigger period +
Max time displayed
Enable: 100 ns
Trigger Impedance
Nominal Impedance 50Ω
Reflection 10% for 100 ps rise time
Connector Type 3.5 mm (male)
Maximum Trigger Signal ±2 V + 2 Vac peak (+16 dBm)
Precision Time Base 86107A
86107A option 010 86107A option 020 86107A option 040
Trigger Bandwidth 2.4 – 4.0 GHz 9.0 – 12.6 GHz 9.0 – 12.6 GHz
9.0 – 12.6 GHz 18.0 – 25.0 GHz 18.0 – 25.0 GHz
39.0 – 43.0 GHz
Typical Jitter (RMS) 2.4 – 4.0 GHz trigger: <280 fs <200 fs 9 – 12.6 GHz, 18 – 25 GHz
trigger bands: <250 fs
9 – 12.6 GHz trigger: <200 fs 38 – 45 GHz trigger: <200 fs
Time Base Linearity Error < 100 fs
Input Signal Type Synchronous clock, no constraint on waveform shape.
Input Signal Level 0.5 – 1.0 Vpp
1
0.2 – 1.5 Vpp (Typical functional performance)
DC Offset Range ±200 mV
Required Trigger Signal-to-Noise Ratio ≥ 200 : 1
Trigger Gating Disable: 0 to 0.6 V
Gating Input Levels (TTL compatible) Enable: 3.5 to 5 V
Pulse width >500 ns, period >1 µs
Trigger Impedance 50 Ω
Connector Type 3.5 mm (male) 3.5 mm (male)
2.4 mm (male)
1
To achieve characteristic performance:
Requires 86100A or B with 86100A software revision 3.0 or above.
For the 86107A with option 020, the Agilent 11742A (DC Block) is recommended if the DC offset magnitude is greater than 200 mV.
1
The freerun trigger mode internally generates an asynchronous trigger that allows viewing the sampled signal amplitude without an external trigger signal but provides no timing information. Freerun is useful in
troubleshooting external trigger problems.
2
The sampled input signal timing is recreated by using an externally supplied trigger signal that is synchronous with the sampled signal input.
3
The DC to 100 MHz mode is used to minimize the effect of high frequency signals or noise on a low frequency trigger signal.
4
Measured at 2.5 GHz with the triggering level adjusted for optimum trigger.
5
High Sensitivity Hysteresis Mode improves the high frequency trigger sensitivity but is not recommended when using noisy, low frequency signals that may result in false triggers without normal hysteresis enabled.