Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
Figure 6-7 Standard Connections for Enhanced Response Calibration
8. To measure the standard, when the displayed trace has settled, press:
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the softkey after it measures the
standard.
9. Disconnect the open, and connect a short circuit to the test port.
10.To measure the device, when the displayed trace has settled, press:
The analyzer measures the short circuit and underlines the softkey.
11.Disconnect the short, and connect an impedance-matched load to the test port.
12.To measure the standard, when the displayed trace has settled, press:
, select the type of load you are using, and then press when
the analyzer has finished measuring the load.
Notice that the softkey is now underlined.
13.To compute the reflection correction coefficients, press:
14.To start the transmission portion of the correction, press .
15.Make a "thru" connection between the points where you will connect your device under
test as shown in Figure 6-7.
OPEN
OPEN
SHORT
SHORT
LOADS
DONE: LOADS
LOADS
STANDARDS DONE
TRANSMISSION