Agilent Technologies 24A Podręcznik Użytkownika Strona 333

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 453
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 332
6-45
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Matched Adapters
With this method, you use two precision matched adapters which are "equal." To be equal,
the adapters must have the same match, Z
0
, insertion loss, and electrical delay. The
adapters in most Agilent calibration kits have matched electrical length, even if the
physical lengths appear different.
Figure 6-17 Calibrating for Noninsertable Devices
To use this method, refer to Figure 6-17 and perform the following steps:
1. Perform a transmission calibration using the first adapter.
2. Remove adapter A, and place adapter B on port 2. Adapter B becomes the effective test
port.
3. Perform a reflection calibration.
4. Measure the test device with adapter B in place.
The errors remaining after calibration with this method are equal to the differences
between the two adapters that are used.
Przeglądanie stron 332
1 2 ... 328 329 330 331 332 333 334 335 336 337 338 ... 452 453

Komentarze do niniejszej Instrukcji

Brak uwag