Agilent Technologies Weather Radio 1680 Instrukcja Użytkownika Strona 69

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 68
69
Chapter 3: Testing Performance
To test the time interval accuracy
b In the pop-up menu, select Rising Edge.
Acquire and verify the test data
1 Click the Run icon to fill acquisition memory.
2 Set up the M1 marker for time interval measurement:
a From the main menu, choose Markers>Properties....
b In the Marker Properties tab of the Waveform Properties dialog, select
the M1 marker.
c In the Position box, select Value.
d Click Occurs....
e In the Value dialog, enter “1” in the Find occurrences field.
Przeglądanie stron 68
1 2 ... 64 65 66 67 68 69 70 71 72 73 74 ... 173 174

Komentarze do niniejszej Instrukcji

Brak uwag