Agilent Technologies Weather Radio 1680 Instrukcja Użytkownika Strona 48

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 47
48
Chapter 3: Testing Performance
To test the multiple-clock state acquisition
To test the multiple-clock state acquisition
Testing the multiple-clock, state acquisition verifies the performance of the
following specifications:
Minimum master-to-master clock time.
Maximum state acquisition speed.
Setup/Hold time for multiple-clock, state acquisition.
This test checks two combinations of data using multiple clocks at two selected
setup/hold times.
Equipment Required
Set up the equipment
1 If you have not already done so, do the following procedures:
“To set up the test equipment and the logic analyzer” on page 23.
“To set up the logic analyzer for the state mode tests” on page 33.
2 Increase the pulse generator channel 2 width to 3.000 ns.
Equipment Critical Specifications Recommended Model/Part
Pulse Generator 200 MHz 3.0 ns pulse width,
< 600 ps rise time
8133A option 003
Digitizing Oscilloscope 6 GHz bandwidth, < 58 ps rise time 54750A w/ 54751A
Adapter SMA(m)-BNC(f) 1250-1200
SMA Coax Cable (Qty 3) 18 GHz bandwidth 8120-4948
Coupler BNC(m)(m) 1250-0216
BNC Test Connector,
6x2 (Qty 4)
Przeglądanie stron 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 173 174

Komentarze do niniejszej Instrukcji

Brak uwag