Agilent Technologies Weather Radio 1680 Instrukcja Użytkownika Strona 33

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 32
33
Chapter 3: Testing Performance
To set up the logic analyzer for the state mode tests
To set up the logic analyzer for the state mode tests
1 Set up the logic analyzer:
a If you have not already done so, do the procedure “To set up the test
equipment and the logic analyzer” on page 23.
b Exit and restart the Agilent Logic Analyzer application to reinitialize
the logic analyzer.
2 Configure the Analyzer Setup dialog:
a Click the Sampling Setup icon.
b Select State - Synchronous Sampling.
c Configure Trigger Position - 100% poststore.
d Select an Acquisition Depth of 8K.
e Click the Buses/Signals tab.
f Click the threshold field of one of the pods. The Threshold Settings
dialog appears.
g In the Threshold Settings dialog, select Standard and ECL (–1.30 V).
h Click OK to close the Threshold Settings dialog.
i Click OK to close the the Analyzer Setup dialog.
Przeglądanie stron 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 173 174

Komentarze do niniejszej Instrukcji

Brak uwag