Agilent Technologies 86100B Instrukcja Obsługi Strona 90

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 166
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 89
4-32
Replacing Instrument Assemblies
To Remove the A13 Acquisition (Option 001)
To Remove the A13 Acquisition (Option 001)
1 Follow the instructions for removing the mainframe cover, on page 4-4.
2 Use a 5/16 inch wrench to remove the three cables labeled: W30, W31, and W32, shown in Figure
4-48.
Figure 4-48. Removing the Cables from the Coaxial Switch
3 Turn the instrument on its side so that you can access the screws that secure the A13 Acquisition
board, then remove the two screws, as shown in Figure 4-49.
Figure 4-49. Removing the A13 Acquisition Board Screws
W30
W3
W31
Przeglądanie stron 89
1 2 ... 85 86 87 88 89 90 91 92 93 94 95 ... 165 166

Komentarze do niniejszej Instrukcji

Brak uwag