Replacing Instrument Assemblies
To Remove the Mainframe Cover
Figure 4-4. Removing the Side Handles
6 To slide the cover off the mainframe, first turn the mainframe upside down on the bench. Place
your hands on each side of the cover, and using your thumbs, push the instrument out the front of
the cover.
7 Once the mainframe has begun to slide forward, you can then set the instrument on its side (see
Figure 4-5) and slide the cover off completely.
Figure 4-5. Sliding Cover off Mainframe