Agilent Technologies 86100B Instrukcja Obsługi Strona 81

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 166
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 80
4-23
Replacing Instrument Assemblies
To Remove the Front Panel Trigger Input
To Remove the Front Panel Trigger Input
CAUTION Electrostatic discharge (ESD) can damage or destroy electrostatic components. All work on
electronic assemblies should be performed at a static-safe work station. See ”Electrostatic
Discharge Information” on page 1-9 for more information on preventing ESD.
1 Disconnect the power cord from the instrument.
WA R N I N G Opening covers or removing parts is likely to expose dangerous voltages. Disconnect the
instrument from all voltages before it is opened.
2 Remove the trim strips from both sides of the front panel, as shown in Figure 4-31.
Figure 4-31. Remove the Trim Strips
3 Use the T-15 TORX driver to remove the four screws (two on each side) that secure the front panel
to the mainframe, as shown in Figure 4-32.
Przeglądanie stron 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 165 166

Komentarze do niniejszej Instrukcji

Brak uwag