Agilent Technologies DisplayPort Instrukcja Użytkownika Strona 34

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 56
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 33
Agilent Technologies, Inc.
DisplayPort Test Implementation Group 34 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
ii. For HBR:
Clock Recovery Method = Second Order PLL
Nominal Bit Rate = 2.7Gb/s
Loop Bandwidth = 10MHz
Damping Factor = 1.51
iii. For RBR use external trigger for ISI and eye height measurements, see C
appendix.
For RBR RJ measurements use:
Clock Recovery Method = Second Order PLL
Nominal Bit Rate = 1.62Gb/s
Loop Bandwidth = 10MHz
Damping Factor = 1.51
g. In Acquisition, select Manual and increase the Record Length to 8.4Mpts or higher. This will
ensure one Million Unit Intervals as specified in the PHY CTS 1.2b. It also represents
repeatable jitter measurements on the CP2520 pattern used for HBR2.
h. Click Calibration, using the defaults.
i. Press Finish from the Calibration screen.
j. Before leaving the Jitter menu, select mUI as the Units.