Figure 18: HBR2 crosstalk Setup for Trigger/Reference Clock Output
e. In the J-BERT N4903B go to PG Setup > Clock/Jitter Output menu.
i. Set the Trigger Output to the following based on the data rate being testing:
Figure 19: Controlling Crosstalk Amplitude on the J-BERT N4903B
Set crosstalk amplitude to
414mV
diff
.
Set cross talk amplitude to
1.35V
diff
.
Set the Clock/Trigger output
amplitude to 0V during this
step. Crosstalk amplitude will
be applied in 9.4.2.
3. If the Sink DUT supports FAUX, ensure that the DUT is transmitting FAUX traffic to emulate
crosstalk during calibration.