The Battery
A-4 Setting Up Your Cardiograph
at least 16 hours is recommended. To charge the battery, plug the cardiograph
into the wall outlet with the switch set to Standby (off).
Figure A-2: Removing the Battery Door.
A. Battery Door
Removing the Battery
To Remove the Battery:
1. Make sure the cardiograph is unplugged from AC power.
2. Turn the cardiograph bottom-side up.
3. Slide the battery door in the direction of the arrow shown in Figure A-2
until it unlatches (approximately 1/2 inch). Lift off the door.
4. Unplug the battery connector from the cardiograph by squeezing the
edges of the connector and pulling it straight out.
5. Remove the battery and cable.
6. If the battery has been removed for storage, replace the battery cover by
placing the battery door into its slots and sliding the door in the opposite
direction of the arrow shown in Figure A-2 until it latches.