Agilent Technologies ESG Dokumentacja Strona 72

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 170
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 71
Performance Tests ESG Family Signal Generators
5. AM Frequency Response
4-14 Calibration Guide
5. AM Frequency Response
This automated test verifies the AM frequency response specifications. The equipment
measures the variations in modulation depth due to changes in the applied AM rate; dc to
10 kHz. The response is relative to a 1 kHz rate set at the test depth, and expressed in dB.
Each frequency is tested with the following sequence:
1. A reference signal is measured.
2. The set depth is measured for each applied rate change.
3. Each depth measurement is compared to the reference signal depth.
The comparison is a calculation of the difference in the depths relative to the reference
rate. The relative value is the AM Frequency Response.
Recommended Equipment
HP/Agilent 8902A Measuring Receiver Option 010
HP/Agilent 8904A Function Generator
HP/Agilent 8663A Signal Generator
MD/MDC-164 Mixer (0.5 to 9 GHz)
Equipment Setups
Figure 4-12 Frequency Response for AM 1300 MHz Test Setup
Przeglądanie stron 71
1 2 ... 67 68 69 70 71 72 73 74 75 76 77 ... 169 170

Komentarze do niniejszej Instrukcji

Brak uwag