Adjustments ESG Family Signal Generators
Level Meter Calibration
5-38 Calibration Guide
Level Meter Calibration
Description
This test determines the level meter gain and offset calibration constants on the output
board. The level meter function is used when the ALC loop is open. It allows the output
power of the signal generator to be monitored and controlled without using the ALC loop.
After the level meter gain and offsets are measured, a calibration constant is calculated
and stored in the appropriate calibration array.
Required Test Equipment
None
Procedure
1. Preset the signal generator.
2. Follow the instructions as they appear on the controller’s display.
Przeglądanie stron 143