Agilent Technologies ESG Dokumentacja Strona 130

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 170
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 129
Adjustments ESG Family Signal Generators
DCFM Calibration
5-24 Calibration Guide
DCFM Calibration
Description
This test removes all of the dc offsets associated with the FM path while in DCFM mode.
This test uses only the FM1 path to verify the functionality of the circuitry. The resulting
values are stored in the signal generator’s firmware.
Required Test Equipment
HP/Agilent 53132A Option 050 Universal Counter
Procedure
Figure 5-20 DCFM Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.
Przeglądanie stron 129
1 2 ... 125 126 127 128 129 130 131 132 133 134 135 ... 169 170

Komentarze do niniejszej Instrukcji

Brak uwag