Eye-Diagram Analyzer Reference
Agilent 70820A User-Corrections
7 Set the frequency scale of the user-correction trace:
If the microwave transition analyzer is set for a frequency sweep, use the
Main menus START and STOP softkeys.
If the microwave transition analyzer is set for a time sweep, press the Main
menus SEC/DIV softkey, and enter the sweep time to set the stop frequen-
cy. The formula to determine the sweep time is provided at the beginning of
this section.
8 To enter the Scale menu, press:
Scale
9 Press the REF LEV|POS softkey, and set the reference level to correctly
position the trace on screen.
10 Change the vertical amplitude scale, if needed, by pressing:
SCALE, Calib, user corr
The user correction data is now displayed graphically above the data.