Agilent Technologies E6392B Dokumentacja Strona 72

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 197
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 71
Chapter 3 71
Programming Command Reference
TESTs Subsystem for AUTOMATIC TEST
3. Programming Command
Reference
DC Current results (TCH1) - Refer to “DC Current Measurement” on page 74
for the meanings of each <string>.
<string 32>, <string 33>, <string 34>, <string 35>, <string 36>:
Peak TX Power results (TCH2)
. . . . .
. . . . .
<string 183>, <string 184>, <string 185>, <string 186>:
DC Current results (TCH6)
All Tests Results (Summary)
TESTs:AUTO:RESults:ALL?
This query command returns a set of values, <string 1> and <string 2>,
that summarize the test results.
<string 1> is the 8-bit binary coded decimal ranging from 0 to 255 according to
the pass/fail results for all test flow steps. Each 8-bit binary coded decimal is
defined as follows:
bit 0 (1): LOC. Update
bit 1 (2): 1st Call
bit 2 (4): Talk
bit 3 (8): RF Test
bit 4 (16): 1st Release
bit 5 (32): 2nd Call
bit 6 (64): 2nd Release
bit 7 (128): Measurement aborted
The choices for each bit are 1 for Fail and 0 for Pass or Skip.
<string 2> is a 10-bit binary coded decimal ranging from 0 to 1023 according to
the pass/fail results for all Test Items in the table. Each 10-bit binary coded
decimal is defined as follows:
bit 0 (1): Peak TX Power
bit 1 (2): Burst Timing
bit 2 (4): Power Ramp
bit 3 (8): Phase Error
bit 4 (16): Frequency Error
bit 5 (32): Sensitivity
bit 6 (64): RX Quality
Przeglądanie stron 71