116 Chapter 3
Programming Command Reference
TRIGger Subsystem
TRIGger Subsystem
The Trigger commands are used to control a measurement cycle.
TRIGger Subsystem Command Reference
Abort
TRIGger:ABORt
This command stops and resets the trigger system, and makes the trigger sequence
state idle. There is no query form of this command.
Trigger Immediately
TRIGger:IMMediate
This command immediately starts a measurement cycle. If one cycle is complete,
the Test Set’s state becomes idle. There is no query form of this command.
Trigger Mode
TRIGger:MODE SINGle|CONTinuous
TRIGger:MODE?
This command sets the trigger mode to either a single or repetitive (continuous)
measurement. The choices are SINGle and CONTinuous. At preset, this is set to
CONT.