Agilent Technologies 1100 Series Przewodnik Instalacji Strona 91

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 184
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 90
1100 LC Modules Control Manual - 91 - 03-914947-30:Rev.6
Automatic Purge Parameter Group
In the Purge time field, enter the time during which the
instrument is purged.
In the Wait time field, enter the time to wait after a purge.
Peaks Parameter Group
Select the Peakwidth value in the scrolling list. This parameter
influences the peak detection. A response time is associated
with each peakwidth value. The available values are:
<0.005 min (RiseTime<0.12 sec)
>0.005 min (RiseTime>0.12 sec)
>0.010 min (RiseTime>0.25 sec)
>0.025 min (RiseTime>0.5 sec)
>0.05 min (RiseTime>1 sec)
>0.1 min (RiseTime>2 sec)
>0.2 min (RiseTime>4 sec)
>0.4 min (RiseTime>8 sec)
Select the Data rate value in the scrolling list. This parameter is
associated to the previous ones. The available values are:
36.50 Pt/sec
18.25 Pt/sec
9.12 Pt/sec
4.56 Pt/sec)
2.28 Pt/sec
1.14 Pt/sec
Przeglądanie stron 90
1 2 ... 86 87 88 89 90 91 92 93 94 95 96 ... 183 184

Komentarze do niniejszej Instrukcji

Brak uwag