1100 LC Modules Control Manual - 83 - 03-914947-30:Rev.6
Settings Control
Mode Parameter Group
Choose the Acquisition Mode between Fluorescence,
Phosphorescence, or Chemo-luminescence.
To work in emission or excitation Multi-wavelength mode, select
the corresponding mode in the scrolling list. When selecting the
excitation multi-wavelength mode, the wavelengths from B to D in
the emission part are not accessible and vice versa.
Select in the Peakdwidth scrolling list, the peakdwidth of the
detector influencing the peak detection. A response time and a
data rate are associated to each peakwidth values.