52 7000 Series Oscilloscopes Service Guide
2 Testing Performance
1 Connect the equipment (see Figure 7).
a Use the N cable to connect the signal generator to the
power splitter input.
b Connect one output of the power splitter to the Ext Trigger
input.
c Connect the power sensor to the other output of the power
splitter.
Figure 7 Setting Up Equipment for 2-Channel External Trigger Sensitiv-
ity Test
Oscilloscope
Signal
Generator
Power Meter
Power Splitter
Power Sensor
N to BNC adapter
N Cable
50 Feedthrough
7000 service guide.book Page 52 Thursday, October 16, 2008 12:18 PM