Chapter 5 67
Mode Parameters
List Sequencer Global Parameters
Mode Parameters
Input Trigger - External Port 2 - Level
This parameter defines the voltage level at the Trigger 2 In port at which triggering occurs.
The remote command for this parameter is:
:TRIGger:LSEQuencer:EXTernal2:LEVel <level>
Trigger Output Port 1 - Polarity
This parameter defines the polarity of the trigger at the Trigger 1 Out port. For a rising edge trigger, set
the polarity to Positive. For a falling edge trigger, set the polarity to Negative.
The remote command for this parameter is:
:TRIGger1[:SEQuence]:OUTPut:POLarity POSitive|NEGative
Trigger Output Port 2 - Polarity
This parameter defines the polarity of the trigger at the Trigger 2 Out port. For a rising edge trigger, set
the polarity to Positive. For a falling edge trigger, set the polarity to Negative.
The remote command for this parameter is:
:TRIGger2[:SEQuence]:OUTPut:POLarity POSitive|NEGative
IF Gain
This parameter allows a 10 dB IF amplifier to be switched on (High Gain) or off (Low Gain), in order to
take advantage of the RF dynamic range of the analyzer.
The remote command for this parameter is:
[:SENSe]:LSEQuencer:IF:GAIN[:STATe] LOW|HIGH