Setting the Logical
Address Switch
Figure 2-1 shows the location and settings of the multimeter’s logical
address switch.
The switch has a factory setting of 24 which is equivalent to a secondary
HP-IB address of 03. If you have more than one multimeter, you must change
the logical address to some other multiple of 8 (for example, 32, 40, 48...), as
there can only be one instrument per secondary address.
Figure 2-1. HP E1326/1411 Logical Address Switch Settings
22 Configuring the HP E1326B/E1411B Multimeter Chapter 2