Agilent E1326B/E1411B 5 1/2 Digit Multimeter User’s Manual E1326-90009 Printed in USA July 2004 E0704 *E132

Strona 2 - Contents

SA DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Revision: A.03 Issue Date: 09/05/00 Manufacturer’s Name:

Strona 3

Enabling OffsetCompensationThe MEASure and CONFigure commands turn offset compensation off.The command used to turn offset compensation on is:RESistan

Strona 4

Figure 4-1. The Multimeter Trigger System102 Understanding the HP E1326B/E1411B Multimeter Chapter 4

Strona 5

The Trigger Source The trigger source parameter specifies the signal which triggers themultimeter. The trigger source is set with the following comma

Strona 6

The Trigger Count The function of the trigger count parameter depends on whether thestand-alone multimeter or scanning multimeter is used.Stand-alone

Strona 7

Example 1: Setting theTrigger Count(stand-alonemultimeter)In this example, one DC voltage measurement is taken each time an externaltrigger occurs. A

Strona 8

The Trigger Delay The trigger delay parameter allows you to specify the period between thetrigger signal and the measurement. For the stand-alone mul

Strona 9 - Safety Symbols

The following program segment shows the context in whichTRIGger:DELay is used.Example: Setting aTrigger Delay 10 !Configure the scanning multimeter fo

Strona 10 - DECLARATION OF CONFORMITY

The Sample Count The sample count specifies the number of measurements made for eachtrigger signal received. For the stand-alone multimeter, it is th

Strona 11

The Sample Period Sample period is the time between measurements in a multiple-readingburst, or the time between channels when scanning a FET multiple

Strona 12 - Multimeter

Table 4-6. Aperture Times and Minimum Sample PeriodAperture Time Minimum Sample Rate(SAMPle:TIMer)Maximum Reading Rate(Readings/second)10 µs76 µs13,15

Strona 13 - Electrical

SA DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Revision: A.03 Issue Date: 09/05/00 Manufacturer’s Name: A

Strona 14 - Physical

TheWait-For-TriggerStateFor the multimeter to respond to a trigger signal, the multimeter must beplaced in the wait-for-trigger state. This is done w

Strona 15 - Introduction to Operation

Using a SingleTriggerThe multimeter can be internally triggered with a single trigger signal. Thissignal is issued with the TRIGger[:IMMediate] comma

Strona 16 - Resetting the

Example: Aborting aMeasurement (Trigger Source = BUS)After the multimeter is configured it is placed in the wait-for-trigger state.ABORt returns the m

Strona 17

Saving Multimeter ConfigurationsTo minimize repeated programming, up to 10 stand-alone multimeterconfigurations can be saved in mainframe/command modu

Strona 18

Example: Saving andRecalling aConfigurationThis program saves a configuration in register 0. The multimeter is thenreset in order to change the curre

Strona 19 - Measurement

Notes116 Understanding the HP E1326B/E1411B Multimeter Chapter 4

Strona 20

Chapter 5HP E1326B/E1411B Multimeter CommandReferenceUsing This ChapterThis chapter describes the Standard Commands for ProgrammableInstruments (SCPI

Strona 21 - Address Switch

CALibration:ZERO:AUTO?Colons separate the root command from the second level command(CALibration:ZERO) and the second level from the third level(ZERO:

Strona 22 - Forming a Scanning

Parameters Parameter Types. The following table contains explanations and examplesof parameter types you might see later in this chapter.Parameter Ty

Strona 23 - VXIbus Interrupt

Table 5-1. Voltage and Ohms Ranges vs. Resolution, Aperture, and Integration TimeAperture Time Integration Time (PLCs) 10 µs*0.0005100 µs0.0052.5 ms0.

Strona 24 - HP E1326B Adapter

Chapter 1Getting Started with the HP E1326B/E1411BMultimeterAbout This ChapterThis chapter introduces you the B-size HP E1326B and C-size HP E1411B51

Strona 25 - Mainframe

SCPI Command ReferenceThis section describes the Standard Commands for ProgrammableInstruments (SCPI) for the HP E1326B and HP E1411B 51⁄2-DigitMultim

Strona 26 - Reference Frequency

ABORtThe ABORt command subsystem removes the multimeter from the wait-for-triggerstate and places it in the idle state. ABORt can only be used with t

Strona 27 - Input Characteristics

CALibrationThe CALibration command subsystem selects the multimeter’s line referencefrequency (CALibration:LFRequency) and enables/disables the autoze

Strona 28

Example Querying the Line Reference FrequencyCAL:LFR 50 !Reference frequency is 50 Hz.CAL:LFR? !Query for reference frequency.enter statement !Enter v

Strona 29 - Multiplexers

Example Querying the Autozero Mode CAL:ZERO:AUTO OFF !Disable autozero.CAL:ZERO:AUTO? !Query multimeter to return autozero mode("0").enter s

Strona 30 - Connections at the

CONFigureThe CONFigure command subsystem configures the multimeter to perform thespecified measurement with the given range and resolution. CONFigure

Strona 31 - Connecting Input Signals

:FRESistanceCONFigure:FRESistance [<range>[,<resolution>]] [,<channel_list>] selects the4-wire ohms function and allows you to spec

Strona 32 - Connections

• Related Commands: FETCh?, INITiate, READ?Example Making 4-Wire Ohms MeasurementsCONF:FRES 1560,MAX,(@100:103) !Function: 4-wire ohms; range selecte

Strona 33 - E1345A/47A/51A/53A

• The channel list is of the form (@ccnn), (@ccnn,ccnn), or (@ccnn:ccnn), wherecc = card number and nn = channel number (105 is channel 05 of card num

Strona 34 - See Chapter 2 of the

<type>discretenumericnumericTC: B, E, J, K, N14, N28, R, S, or TTHER/FTH:2252 | 5000 | 10000RTD/FRTD: 85 | 92noneohmsalpha<channel_list>nu

Strona 35 - E1344A/47A/53A

set the addresses in order to create an instrument. The guide should be yourstarting point toward using the multimeter. The functions and features o

Strona 36 - Carrier Cable Assemblies

<resolution>numericresolution (see Table 5-1) |DEF | MIN | MAXvolts<channel_list>numeric Refer to the multiplexer user’s manual fora list

Strona 37

:VOLTage[:DC]CONFigure:VOLTage[:DC] [<range>[,<resolution>]] [,<channel_list>] selects theDC voltage function and allows you to spe

Strona 38 - Additional Configurations

parameter). This prevents the MIN or MAX resolution from being interpreted as arange setting.• The fastest aperture time available when autoranging i

Strona 39

CONFigure?The CONFigure? command queries the multimeter to return the configuration set bythe most recent CONFigure or MEASure command. It returns a q

Strona 40 - About This Chapter

DIAGnosticThe DIAGnostic command subsystem provides control of the FET multiplexers.Subsystem Syntax DIAGnostic:FETS <mode>:FETS?:FETSDIAGnostic

Strona 41 - Making a Single Measurement

DISPlayThe DISPlay command subsystem monitors the state of the selected multiplexerchannel within the scanning multimeter. This command is useful onl

Strona 42

Example Monitoring a ChannelDISP:MON:CHAN (@101) !Select channel 101 for monitor mode.DISP:MON ON !Enable monitor mode.:MONitor:CHANnel?DISPlay:MONit

Strona 43

:MONitor[:STATe]?DISPlay:MONitor[:STATe]? returns a number to show whether the monitor modeis enabled or disabled: "1" = ON, "0"

Strona 44

FETCh?The FETCh? command retrieves measurements stored in mainframe/command module memory by the most recent INITiate command and places them inthe ou

Strona 45 - Scanning a Channel List

FORMatThe FORMat subsystem sets the format for data transferred from the multimeter tothe computer using the MEASure?, READ?, and FETCh? commands.Subs

Strona 46 - Making Multiple Scans

Table 1-1. HP E1326B/E1411B Operating CharacteristicsDC VoltageRangesResolutionAccuracy (90 days)Max Rdgs/sec0.125V, 1.0V, 8.0V, 64.0V, 300V full scal

Strona 47 - Making Multiple Paced Scans

FORMat?FORMat? returns one of the following to the output buffer:• ASC,+7 seven significant digits• REAL,+32 REAL 32 format• REAL,+64 REAL 64 format1

Strona 48

INITiateThe INITiate command subsystem places the multimeter in the wait-for-trigger state.This command is most commonly used with CONFigure.Subsystem

Strona 49

MEASureThe MEASure command subsystem configures the multimeter to perform thespecified measurement with the given range and resolution. When the mult

Strona 50

:FRESistance?MEASure:FRESistance? [<range>[,<resolution>]] [,<channel_list>] selects the4-wire ohms function and allows you to speci

Strona 51

Example Making 4-Wire Ohms MeasurementsMEAS:FRES? 1560,MAX,(@100:103) !Function: 4-wire ohms;range selected: 1861 Ω;MAX resolution: 125 mΩ;specify sen

Strona 52

• To specify a MIN or MAX resolution while autoranging, you must specifyMEAS:RES? AUTO or MEAS:RES? DEF must be specified (you cannot omit therange pa

Strona 53 - Multiple High-Speed Scans

where “cc” is the multiplexer (card) number.• The channel list is of the form (@ccnn), (@ccnn,ccnn), or (@ccnn:ccnn), wherecc = card number and nn = c

Strona 54

• The MIN and MAX parameters select the minimum or maximum values for rangeand resolution:For range: MIN = 0.0795 V; MAX = 300 V.For resolution: MIN s

Strona 55 - Maximizing Measurement Speed

Comments • The [:DC] parameter is optional. Both of the following command statementsselect the DC voltage function:MEAS:VOLT:DC? or MEAS:VOLT?

Strona 56

MEMoryThe MEMory command subsystem enables you to store multimeter readings inshared memory (an external VME memory card).Subsystem Syntax MEMory:VME:

Strona 57 - Changing the Data Format

Introduction to OperationThis section contains information on checking communication between themultimeter, mainframe, and computer. It includes info

Strona 58 - ENTER when all ENTER

The lowest decimal address available (2097152) if MIN is specified.The highest decimal address available (14680060) if MAX is specified.Example Queryi

Strona 59 - HP 82335 HP-IB Interface Card

Example Querying the VME Memory SizeMEM:VME:SIZE 100000 !Set memory size to 100 kBytes.MEM:VME:SIZE? !Query multimeter to return memory size.enter sta

Strona 60

OUTPutThe OUTPut command subsystem enables you to route the multimeter’s voltmetercomplete signal to the VXIbus TTL trigger lines.Subsystem Syntax OUT

Strona 61

Example Routing Voltmeter Complete to Trigger LineOUTP:TTLT7 ON !Route signal to trigger line 7.:TTLTrgn[:STATe]?OUTPut:TTLTrgn[:STATe]? returns a nu

Strona 62

READ?The READ? command is most commonly used with CONFigure to:Place the multimeter in the wait-for-trigger state (executes the INITiate command).Tran

Strona 63

READ? !Place multimeter in wait-for-trigger state andmake measurements; send readings to outputbuffer; trigger source is IMMediate bydefault.enter st

Strona 64 - MEM:VME:STAT ON), all

SAMPleThe SAMPle command subsystem operates with the TRIGger command subsystem.The SAMPle subsystem:Designates the number of readings made for each tr

Strona 65 - Checking for Errors

READ? !Place multimeter in wait-for-trigger state;make measurement when external trigger isreceived; send readings to output buffer.enter statement !E

Strona 66

READ? !Place multimeter in wait-for-trigger state andmake measurements; send readings to outputbuffer.enter statement !Enter readings into computer.:S

Strona 67

• Related Commands: SAMPle:COUNt, SAMPle:SOURce, SENSe•*RST Condition: SAMPle:TIMer 50E-3 secondsExample Setting the Sample PeriodCONF:VOLT:DC !Func

Strona 68 - Additional Scanning

Table 1-2. HP E1326/E1411 Self-Test CodesSelf-Test Code Description0 Test passed.1 Multimeter does not respond to the self-test.2 Invalid communicatio

Strona 69

[SENSe:]The [SENSe:] command subsystem is most commonly used with CONFigure tochange specific "low-level" measurement parameters. Normally

Strona 70 - Subprogram Kaboom

:RESolution <resolution>:RESolution? [MIN | MAX]Note The root command [SENSe:] is an implied command and can be omitted.FUNCtion[SENSe:]FUNCtion

Strona 71

RESistance:APERture[SENSe:]RESistance:APERture <time> sets the aperture (integration time) inseconds. Values are rounded up to the nearest ape

Strona 72 - Get readings

RESistance:APERture?[SENSe:]RESistance:APERture? [MIN | MAX] returns one of the followingnumbers to the output buffer:The present aperture time in se

Strona 73

RESistance:NPLC?[SENSe:]RESistance:NPLC? [MIN | MAX] returns one of the following numbers tothe output buffer:The present integration time in PLCs if

Strona 74

Example Querying the Offset Compensation ModeRES:OCOM ON !Enable offset compensation.RES:OCOM? !Query multimeter to return offsetcompensation mode.ent

Strona 75

RESistance:RANGe?[SENSe:]RESistance:RANGe? [MIN | MAX] returns one of the following numbersto the output buffer:The present resistance range is selec

Strona 76 - RESistance:OCOMpensated

RESistance:RANGe:AUTO?[SENSe:]RESistance:RANGe:AUTO? returns a number to show whether theautorange mode is enabled or disabled: "1" = ON, &

Strona 77 - How to Make Measurements

Example Changing the ResolutionCONF:FRES 1560,MAX,(@100:103) !Function: 4-wire ohms;range selected: 1861 Ω;MAX resolution: 125 mΩ;specify sense channe

Strona 78 - When Using CONFigure

•MIN selects the minimum range available with the VOLTage:AC:RANGecommand: 0.0875V. MAX selects the maximum range available: 300 V.• You must select

Strona 79 - Data Formats and Destinations

Table 1-3. HP E1326/E1411 Power-on SettingsParameter SettingFUNCtion VOLT:DCVOLTage:RANGe 8VRESistance:RANGe16384ΩVOLTage:RANGe:AUTO ONRESistance:RANG

Strona 80 - Destinations

VOLT:AC:RANG? !Query multimeter to return the presentrange.enter statement !Enter value into computer.VOLTage:APERture[SENSe:]VOLTage:APERture <tim

Strona 81

Example Querying the Aperture TimeVOLT:APER 2.67E-01 !Aperture time is 267 ms.VOLT:APER? !Query multimeter to return aperture time.enter statement !En

Strona 82

VOLT:DC:RANG 0.9 !Range selected 1 V; MAX resolution: 61.035 µV.READ? !Place multimeter in wait-for-trigger state andmake measurements; send readings

Strona 83 - Shared Memory

• The greater the number of PLCs, the greater the normal mode rejection (and thelower the reading rate).• Only the 1 PLC and 16 PLC settings provide n

Strona 84 - Destination

• Related Commands: CONFigure, VOLTage:RANGe•*RST Condition: VOLT:RANG:AUTO ONExample Disabling AutorangingVOLT:RANG:AUTO OFF !Disable autorange.VOL

Strona 85 - Measurement Functions

• The VOLT:RANG command overrides the range setting from a previousCONFigure command on the same function. The multimeter uses the sameaperture time

Strona 86 - Two-Wire vs. Four-Wire

SYSTemThe SYSTem command subsystem returns error numbers and messages in the errorqueue. For the scanning multimeter configuration only, SYSTem can a

Strona 87

• HEWLETT-PACKARD,E1347A,0,A.01.00• HEWLETT-PACKARD,E1460A,0,A.01.00• NONE,NONE,0,0ParametersParameter Name Parameter Type Range of Values Default Uni

Strona 88 - Thermocouple

TRIGgerThe TRIGger command subsystem controls the behavior of the trigger system. Thesubsystem can control:The number of triggers to occur before the

Strona 89 - Function

•CONFigure and MEASure set the trigger count to 1.• *RST Condition: TRIG:COUN 1Examples Setting the Trigger Count (Scanning Multimeter)CONF:VOLT:DC (

Strona 90 - Multimeter Parameters

Clearing the Multimeter When the multimeter is selected from the HP E1301A mainframe menu, themultimeter is cleared by pressing the “Clear Instr” key

Strona 91

:DELayTRIGger:DELay <period> sets the delay period between receipt of the trigger andthe start of the measurement.ParametersParameter Name Para

Strona 92

:DELay:AUTOTRIGger:DELay:AUTO <mode> enables or disables a trigger delay based on thepresent function, range, and integration time. The trigge

Strona 93

[:IMMediate]TRIGger[:IMMediate] causes a trigger to occur immediately provided themultimeter is in the wait-for-trigger state (see the INITiate subsy

Strona 94 - 1.084540E+001

ParametersParameter Name Parameter Type Range of Values Default Units<source>discrete BUS | EXT | HOLD | IMMediate | TTLTrg0 - TTLTrg7noneComme

Strona 95 - Resolution

Example Querying the Trigger SourceTRIG:SOUR EXT !Trigger source is external BNC onmultimeter front panel.TRIG:SOUR? !Query multimeter to return trigg

Strona 96 - Integration Time

IEEE 488.2 Common Command ReferenceThe following table lists the IEEE 488.2 Common (*) Commands that can beexecuted by the HP E1326B and HP E1411B 51

Strona 97 - Querying the Aperture

Command Quick Reference The following tables summarize SCPI commands for the HP E1326B andHP E1411B 51⁄2-Digit Multimeters.Command DescriptionABORt Pl

Strona 98 - Querying the Autozero

Command DescriptionSAMPle :COUNt 1-16777215 | MIN | MAX:COUNt? [MIN | MAX]:SOURce IMM | TIM:SOURce?:TIMer 76 µs-65.534 ms | MIN | MAX:TIMer? [MIN | MA

Strona 99 - Compensation

Appendix A 1326B/E1411B Multimeter Specifications 189 Appendix A E1326B/E1411B Multimeter

Strona 100 - Triggering the Multimeter

DC VoltageAccuracy conditions for table below: Autozero on, one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at

Strona 101

The error queue can store up to 30 error messages which are retrieved in afirst in, first out (FIFO) manner. When there are no error messages in theq

Strona 102 - Querying the Trigger

Appendix A 1326B/E1411B Multimeter Specifications 191 Accuracy conditions for table below: Autozero on, one hour warmup. Within 1 year and ±5°

Strona 103

192 1326B/E1411B Multimeter Specifications Appendix A Four-Wire Resis

Strona 104

Appendix A 1326B/E1411B Multimeter Specifications 193 Accuracy condition for table below: Autozero on, one-hour warmup. Within 90 days and ±5°

Strona 105

194 1326B/E1411B Multimeter Specifications Appendix A True RMS AC Vol

Strona 106 - Trigger Delay

Appendix A 1326B/E1411B Multimeter Specifications 195 Accuracy conditions for table below: Autozero on, one-hour warmup. Within 90 days and ±5

Strona 107 - Querying the Sample

Timer/Pacer Module Size/Device Type:Timer Range: 76 µsec to 65.5 msec B, register-based (E1326B)Resolution: 2 µs C, register-based (E1411B)Accuracy: 0

Strona 108

Appendix BHP E1326B/E1411B Multimeter ErrorMessagesThe error messages associated with the HP E1326B/HP E1411B multimeterare shown in this appendix.Co

Strona 109

Code Message Cause- 230 Data corrupt or stale Data in mainframe memory is fetched after a command(e.g., MEASure, READ?) has returned data to theoutpu

Strona 110 - Wait-For-Trigger

Appendix CHP E1326B/E1411B MultimeterRegister-Based ProgrammingAbout This AppendixThe HP E1326B and HP E1411B 51⁄2-Digit Multimeters are register-bas

Strona 111 - Single Trigger

The Base Address When you are reading or writing to a multimeter register, a hexadecimal ordecimal register address is specified. This address consis

Strona 112

ContentsHP E1326B/E1411B 5 1/2 Digit Multimeter User’s ManualWarranty . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 113 - Configuration

Chapter 2Configuring the HP E1326B/E1411BMultimeterAbout This ChapterThis chapter contains information on connecting input signals to themultimeter u

Strona 114 - Recalling a

Finding the BaseAddress in anEmbedded ControllerWhen using an embedded controller such as the HP RADI-EPC7 with theStandard Instrument Command Library

Strona 115

Register Offset The register offset is the register’s location in the block of 64 address bytes.For example, the multimeter’s Command Register has an

Strona 116 - Reference

Register DescriptionsThere are three WRITE and five READ registers on the multimeter. Thissection contains a description and a bit map of each regist

Strona 117 - Commands

The Command andParameter RegistersCommands and their parameters are opcodes written to the Command andParameter Registers.Address 15 - 8 76543210base

Strona 118

The READ Registers The following READ registers are located on the multimeter.– ID Register (base + 0016)– Device Type Register (base + 0216)– Status

Strona 119

The Device TypeRegisterThe Device Type Register contains a model code which identifies thedevice.Address 1514131211109876543210base + 0216Model CodeMo

Strona 120 - SCPI Command Reference

Passed. A zero (0) in bit 2 indicates the multimeter is executing a reset, oris executing or failed its self-test. A one (1) indicates the reset is

Strona 121 - Comments •

The Data Buffer Measurements are returned to the Data Buffer. The buffer is afirst-in-first-out (FIFO) buffer capable of storing 256 four-byte readin

Strona 122 - CALibration

Converting Four-Byteand Two-Byte ReadingsFour-byte readings and two-byte readings are converted to voltages andresistances as follows.Four-Byte Readin

Strona 123 - :ZERO:AUTO?

Program Timing and ExecutionWhen programming the multimeter at the register level, the structure of theprogram will generally be as follows:– resettin

Strona 124 - Querying the Autozero Mode

Setting the LogicalAddress SwitchFigure 2-1 shows the location and settings of the multimeter’s logicaladdress switch.The switch has a factory setting

Strona 125 - CONFigure

• Bit 2 of the Status Register is monitored to determine when the resetis finished.• Writing a "0" to bit 1 re-enables SYSFAIL. Configuring

Strona 126 - :FRESistance

Figure C-4. Configuring the Multimeter212 HP E1326B/E1411B Multimeter Register-Based Programming Appendix C

Strona 127 - :RESistance

RetrievingMeasurementsFigure C-5 shows the conditions monitored to determine when measurementsare available in the data buffer.Comments• The registers

Strona 128 - :TEMPerature

Checking for Errors Error conditions are monitored and error codes are returned as indicated infigure C-6.Comments• The registers used are:– Status Re

Strona 129 - :VOLTage:AC

QueryingParametersMultimeter parameters are queried as shown in Figure C-7.Comments• The registers used are:– Status Register (base + 0416)– Command R

Strona 130

Using a Multiplexerwith the MultimeterFigure C-8 shows an example timing sequence between closing a multiplexerchannel and triggering the multimeter.C

Strona 131 - :VOLTage[:DC]

Register TriggeringThis section reviews the multimeter’s trigger system from the register-basedstandpoint. The section shows the triggering models us

Strona 132

MultimeterTriggering ModelThe following models suggest one method of triggering the multimeterwhen it is used to make stand-alone or scanning measurem

Strona 133 - CONFigure?

Control RegisterSamplingThe following model shows how to make a measurement by writing to theControl Register. This method of sampling is available w

Strona 134 - DIAGnostic

Programming ExamplesThe examples in this section demonstrate how to program the multimeter atthe register level. The programs follow the execution an

Strona 135 - :MONitor:CHANnel

Forming a ScanningMultimeterIf multiplexers are used to form a scanning multimeter, they must beassigned successive logical addresses beginning with t

Strona 136 - :MONitor[:STATe]

Resetting theMultimeterThe following program resets the multimeter.HP BASIC/WS 10 !Map the A16 address space in the HP V/382 and store the multimeter

Strona 137 - :MONitor[:STATe]?

C Version/* E1411_RS.CPP - This program resets the multimeter. */#include <sicl.h>#include <stdio.h>#include <stdlib.h>#include <

Strona 138 - Subsystem Syntax FETCh?

Reading the IDRegisterAs mentioned previously, the ID Register indicates the classification,addressing mode, and manufacturer of the device. This prog

Strona 139 - Comments

Reading the DeviceType RegisterThe Device Type Register contains the model code of the device. The HP E1326B model code is FF4016 and the HP E1411B m

Strona 140 - REAL,+64 REAL 64 format

C Version/* E1411_ID.CPP - This program reads the multimeter’s ID and Device Type *//* registers. */#include <sicl.h>#include <stdio.h>#in

Strona 141 - INITiate

Reading the QueryResponse RegisterThe following program sets the multimeter function to (2-wire) OHMS, andthen queries the function and reads it from

Strona 142 - Subsystem Syntax MEASure

380 !This subprogram sets and queries the multimeter function.390 SUB Func_qry400 COM Base_addr410 WRITEIO -16,Base_addr+8;4420 CALL Wait_not_bsy430

Strona 143 - :FRESistance?

C Version /* E1411_QY.CPP - This program sets the multimeter function to (4-wire) *//* OHMS and then queries the function setting and reads it from

Strona 144 - :RESistance?

/* this function sets the multimeter function to ohms */unsigned short query; /* variable for data from query response reg *//* write ’measurement fu

Strona 145 - :TEMPerature?

Reading an ErrorCodeThis program generates an error and then reads the error code from theQuery Response Register.HP BASIC/WS Version 10 !Map the A16

Strona 146 - :VOLTage:AC?

VXIbus InterruptLinesThe multimeter sends interrupts to, and receives acknowledgements fromthe slot 0 module via the VXIbus backplane interrupt lines.

Strona 147 - :VOLTage[:DC]?

380 !This subprogram calls Read_status to check status bit 1 to determine390 !when an error code is in the Query Response register.400 SUB Qry_ready41

Strona 148

C Version /* E1411_EC.CPP - This program generates an error and then reads the *//* error code from the Query Response Register. */#include <si

Strona 149 - :VME:ADDRess?

/* this function specifies a parameter opcode of 7 for the function *//* parameter. This is a parameter out of range. *//* write ’measurement function

Strona 150 - :VME:SIZE?

Stand-AloneMultimeterMeasurementsThe following program makes measurements using the stand-alonemultimeter. The configuration shown makes five bursts

Strona 151 - :VME:STATe?

360 !Dimension a computer variable to store the measurements.370 !(size = number of bursts * number of readings per burst.)380 REAL Readings(1:25) 390

Strona 152 - :TTLTrgn[:STATe]

770 !This subprogram calls Read_status to check the validity of status bit 7780 !(DONE). This ensures the validity of status bits 6,5,4, and 1.790 SUB

Strona 153 - :TTLTrgn[:STATe]?

1190 !This subprogram writes commands and parameters to the Command1200 !and Parameter registers. It calls Wait_not_bsy prior to writing the next1210

Strona 154 - Subsystem Syntax READ?

1620 Header=BIT(Lower_word,0)*128+Rng1630 Exp=BINAND(Header,7)1640 ELSE1650 Upper_word=READIO(-16,Base_addr+12)1660 CALL Data_ready1670 Lower_word=REA

Strona 155

C Version /* E1411_M.CPP - This program takes measurements on the front *//* terminals of the E1411B multimeter. */#include <sicl.h>#incl

Strona 156 - Subsystem Syntax SAMPle

/* map the E1411B registers into user memory space */base_addr_dmm = imap(e1411b, I_MAP_VXIDEV, 0, 1, NULL);/* function call to reset the multimeter *

Strona 157

HP E1326B InternalInstallationWhen the HP E1326B is installed in an HP E1300A/E1301A/E1302Amainframe, it occupies one slot. However, the faceplate to

Strona 158 - :SOURce?

/* write command to command register *//* wait for mm ready bit = 1 */while(!READY);iwpoke((unsigned short *)(base_addr_dmm + 0x08),command);/* write

Strona 159

iwpoke((unsigned short *)(base_addr_dmm + 0x0A),parm2);while(!DONE); /* wait until mm is done before sending the *//* next command and parameters */if

Strona 160 - [SENSe:]

/******************************************************/void no_dmm_parameters(char *base_addr_dmm, int command){/* this function sends those multime

Strona 161 - FUNCtion?

/* get lower word of reading from the data register, add the *//* word to the reading upper word in the variable dmm_4byte, *//* ensure upper word of

Strona 162 - RESistance:APERture

dmm_2byte = (dmm_2byte >> 1);/* get range code from rng variable, get range from *//* the appropriate range table */if (func == 2)range = ohm_ra

Strona 163 - RESistance:NPLC

ScanningMultimeterMeasurementsThe following program makes measurements using the multimeter and theHP E1460A 64-Channel Relay Module. The configurati

Strona 164 - RESistance: OCOMpensated?

360 !Set the sample source(IMMEDIATE)370 CALL Peek_meas(7,0)380 !Set the sample rate (Set if sample source is Timer (1))390 !CALL Peek_meas(10,0,70)4

Strona 165 - RESistance:RANGe

790 SUBEND800 !This subprogram calls Read_status to check the validity of bit 7 (DONE).810 !This ensures the validity of bits 6,5,4, and 1.820 SUB Wai

Strona 166 - RESistance:RANGe :AUTO

1210 CALL Read_status(Status,Rdy,Done,Indardy,Qryrdy,Noerr,Mmcomp, Pass_fail)1220 UNTIL Indardy1230 SUBEND1240 !This subprogram monitors bit 5 (MULTIM

Strona 167 - RESistance:RESolution

1630 !This subprogram writes commands and parameters to the Command and1640 !Parameter registers. When a command is written, it calls Wait_not_bsy1650

Strona 168 - VOLTage:AC:RANGe

Installing theHP E1411B in aMainframeThe HP E1411B multimeter can be installed in any slot (except slot 0) in aC-size VXIbus mainframe. Refer to Figu

Strona 169 - Example Changing the Range

2000 !This subprogram reads the measurements (all channels) from the 2010 !data buffer. It converts the measured quantities to volts, ohms, etc.,2020

Strona 170 - VOLTage:APERture?

2340 !This subprogram resets the multimeter by disabling the SYSFAIL bit,2350 !then writing a "1" to Control register bit 0, and then writi

Strona 171 - VOLTage[:DC]:RANGe

C Version This program uses the E1411B multimeter and the E1351A FETmultiplexer to perform high-speed scanning. The program configures themultimeter

Strona 172 - VOLTage:NPLC

char *base_addr_dmm; /* pointer to multimeter base address */char *base_addr_fet; /* pointer to multiplexer base address */clrscr( ); /* clear the use

Strona 173 - VOLTage:RANGe:AUTO

/* Download FET multiplexer scan list */set_scanlist(base_addr_fet);/* trigger the multimeter */ trigger_dmm(base_addr_dmm);/* close the device sessio

Strona 174 - VOLTage:RESolution

/*******************************************************/void set_sample_period(char *base_addr_dmm, int command, unsignedshort parameter){/* this fu

Strona 175 - VOLTage:RESolution?

/******************************************************/void set_dmm_counts(char *base_addr_dmm, int command, long parameter){/* this function sets t

Strona 176 - :CDEScription?

/******************************************************/void no_dmm_parameters(char *base_addr_dmm, int command){/* this function sends those multime

Strona 177

/******************************************************/void trigger_dmm(char *base_addr_dmm){/* this function triggers the multimeter and retrieves a

Strona 178 - Subsystem Syntax TRIGger

/* get range code from reading, get range from *//* the appropriate range table */range_code = ((dmm_4byte >> 24) & 7);if (func == 2)range =

Strona 179

The ReferenceFrequencyIn many data acquisition applications, DC voltage and resistancemeasurements are often made in the presence of normal mode noise

Strona 180

/******************************************************/void check_for_error(char *base_addr_dmm){unsigned short error_code; /* variable for error cod

Strona 181 - :DELay:AUTO

Useful TablesThe tables contained in this section are:• Command and Parameter Opcodes• Register-Based Programming Error Codes• Multimeter Power-On Set

Strona 182 - [:IMMediate]

Table C-2. Command and Parameter Opcodes (continued) MultimeterParameterCommandOpcodeParameterOpcode ValueQueryOpcodeQueryResponseVoltmeterComp

Strona 183

Register-BasedProgramming ErrorCodesThe error codes related to register-based programming are listed in Table C-3.Note When an error occurs, bit 6 in

Strona 184 - Querying the Trigger Source

MultimeterPower-On SettingsThe multimeter’s power-on settings are shown in Table C-4.Table C-4. Multimeter Power-On SettingsParameter Value OpcodeFun

Strona 185 - Synchronization

Function andAperture ChangeTimesTable C-5 lists the times required for the multimeter to change its functionand aperture time. The times pertain to b

Strona 186 - Command Quick Reference

VME Interrupts The HP E1326B/E1411B generates interrupt signals on the conditionsindicated in Table C-6. These signals are available to the computer(

Strona 187 - Select range

Notes268 HP E1326B/E1411B Multimeter Register-Based Programming Appendix C

Strona 188 - Specifications

Appendix DMeasurement Speed and Accuracy TradeoffsThe HP E1326B SCPI driver was designed so that its default mode ofoperation will deliver high accur

Strona 189 - DC Voltage

Avoid FunctionChangesThe E1326B multimeter (DMM) takes time to switch between its variousfunctions because the hardware is reconfigured and calibratio

Strona 190 - (volts)

Input CharacteristicsThe multimeter is a floating, balanced differential multimeter. Floating meansthe multimeter’s input terminals are isolated from

Strona 191 - For Offset

Do Binary Transfers tothe ComputerREAL 32 is fastest and is compatible with compiled languages.REAL 64 is fast and is compatible with HP BASIC.The def

Strona 192

150 GOTO End160 Kaboom: PRINT ""170 PRINT ERRM$180 PRINT "HERE IS THE E13XX ERROR STATUS"190 E13xx_errors200 End: END210 !This s

Strona 193 - µV 30 µV 61 µV

550 ENTER @Dmm;Readings(*)560 Stop=TIMEDATE570 PRINT "TIME TO READ 10000 READINGS IN ASCII=";(Stop-Start)580 !Now transfer the same data in

Strona 194

Use Macros toMinimize CommandParsing TimeMacros provide the fastest programming method when it is necessary tochange functions from one measurement to

Strona 195

Turn Auto Ranging OFF Turning auto ranging OFF makes the E1326B take all of its measurementson a fixed range which results in fast and predictable mea

Strona 196 - Messages

Setting the Resolution The aperture time is set as a result of specifying the <resolution> parameterin the MEASure or CONFigure command, or by d

Strona 197 - Code Message Cause

Store the Readings inCommand ModuleRAM Instead ofSending them Directlyto the ComputerThere is a major difference between INIT;:FETCH? and READ? after

Strona 198 - Register-Based Programming

READ? The READ? command causes the E1326B multimeter to start takingreadings as soon as the trigger requirements have been met (seeINIT;:FETCh).The RE

Strona 199

IndexHP E1326B/E1411B 5 1/2 Digit Multimeter User’s Manual*CLS, 178, 186*DMC, 186*EMC, 186*EMC?, 186*ESE, 186*ESE?, 186*ESR?, 186*GMC?, 186*IDN?, 186

Strona 200

enabling, 167, 174parameters, 94querying setting, 95, 168, 175RESistance:RANGe:AUTO, 94, 167setting, 94VOLTage:RANGe:AUTO, 94, 174AutozeroCALibration:

Strona 201 - Registers

Input Terminals The multimeter input terminals are shown in Figure 2-7. The maximuminput on the HI and LO terminals is 300 V dc (450 V ac peak). The

Strona 202 - Register Descriptions

INITiate, 142linking, 119MEASure, 42, 76 - 78, 82, 143 - 149MEMory, 150 - 152opcodes, 262 - 263OUTPut, 153 - 154parsing time, 274quick reference, 187

Strona 203 - Parameter Registers

REAL-32, 80, 140, 271REAL-64, 58, 80, 140, 271selecting, 80specifying, 80, 140vs. reading destination, 81DC Voltage Measurements, 42accuracy condition

Strona 204 - – Data Buffer (base + 0C

scanning multimeter measurements, 20, 246 - 261setting aperture time and sample period, 110setting sample count, 108setting trigger count, 105setting

Strona 205 - Register

reference thermistor, 89terminal module connections, 33 - 36thermocouple measurements, 130, 146thermocouple types, 89HP E1403B Adapter, 37cables, 38HP

Strona 206 - Response Register

measurement speed, 56 - 57MEASurecommands, 143 - 149making measurements with, 78, 82parameters, 91subsystem, 143 - 149used as a single command, 20used

Strona 207 - Two-Byte Reading

burst measurements, 45high-speed scans, 54 - 55scans, making, 47 - 48MultiplexerAC voltage measurement, 130, 147connected to multimeter, 14, 29 - 31,

Strona 208 - Converting Four-Byte

register-based, 199 - 268timing and execution, 210QQuery Response Register, 207, 226 - 229reading an error code, 230 - 233QueryingAC voltage range, 17

Strona 209 - Program Timing and Execution

error codes, 264examples, 220 - 261function change times, 266ID register, 205, 223parameter register, 204parameters opcodes, 262 - 263power-on setting

Strona 210 - Configuring the

Saving Multimeter Configurations, 114*SAV, 114example program, 115how to, 114in memory, 14ScanningB-size switchbox channels, 50 - 51C-size switchbox c

Strona 211

integration time, 98, 163 - 164, 171, 173IRQ jumper, 24line reference frequency, 123logical address switch, 22 - 23pacing source, 158range, 93, 166rea

Strona 212 - Measurements

ConnectingMultiplexersIn a scanning multimeter configuration, the multimeter is connected to themultiplexers with an analog bus cable, or with the ana

Strona 213 - – Command Register (base + 08

delay, 106 - 107, 181 - 182group execute (GET), 103, 122immediately, 183lines (TTL), 153 - 154model, 218readings per, 157registers, 217 - 219sample co

Strona 214 - • The registers used are:

Notes292 HP E1326B/E1411B 5 1/2 Digit Multimeter User’s Manual Index

Strona 215 - Using a Multiplexer

Making Multiple Burst Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 45Scanning a Channel List . . . .

Strona 216 - Register Triggering

Analog BusConnections at theMultimeterThe analog bus coming from the multiplexer consists of six lines. On themultiplexer terminal block these lines

Strona 217 - Triggering Model

Connecting Input SignalsThis section contains guidelines on connecting input signals to themultimeter and shows the connections required to make the f

Strona 218

MeasurementConnectionsE1460AE1346AE1352AE1345A/47A/51A/53AFigure 2-12. Connections for DC and AC Voltage MeasurementsChapter 2 Configuring the HP E13

Strona 219 - Programming Examples

Note 2-wire resistance measurements require the multiplexer modules shownabove. Resistance measurements using the multimeter terminals or directlythr

Strona 220 - HP BASIC/WS

NOTE: Channel Pairs are banks 0/4, 1/5, 2/6, and 3/7.See Chapter 2 of the HP E1460A User’s Manual.E1460AE1345A/47A/51A/53AFigure 2-14. Connections for

Strona 221 - C Version

E1344A/47A/53AE1476AFigure 2-15. Connections for Thermocouples36 Configuring the HP E1326B/E1411B Multimeter Chapter 2

Strona 222 - HP BASIC/WS Version

Carrier Cable AssembliesThe following table and figures show the cables used to connect relay andFET multiplexers to the HP E1411B multimeter. These

Strona 223

Figure 2-16. Cables for B-Size Multiplexers in HP E1403B Adapter38 Configuring the HP E1326B/E1411B Multimeter Chapter 2

Strona 224

Additional ConfigurationsThis section contains information on two additional configurations for themultimeter:• selecting VME RAM, and• disabling fron

Strona 225

Notes40 Configuring the HP E1326B/E1411B Multimeter Chapter 2

Strona 226

Saving Multimeter Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114How to Save and Recall a Conf

Strona 227

Chapter 3Using the HP E1326B/E1411B MultimeterAbout This ChapterThis chapter is a collection of example programs which show you how tomake measuremen

Strona 228

Multimeter Connections Chapter 2 contains information on connecting input signals for themeasurements described in this chapter.The MEASure andCONFigu

Strona 229

Making a Burst of MeasurementsThis program makes 100 DC voltage measurements on the terminalsconnected to the multimeter’s faceplate. 10 !Dimension a

Strona 230

Making an Externally Triggered Burst of MeasurementsThis program makes a burst of 10 measurements on the faceplate terminalswhen the multimeter receiv

Strona 231

Making Multiple Burst MeasurementsThis program makes multiple burst measurements with a 5 second delaybetween bursts. There are three bursts, each co

Strona 232

• The trigger delay is the period between the trigger signal and thestart of the measurement (burst). The trigger delay set byCONFigure is 0 seconds

Strona 233 - HP BASIC/WS Version

Making Multiple ScansThis program scans a channel list multiple times. 10 !Dimension a computer array to store the readings. 20 DIM Rdgs(1:20) 30 !Cle

Strona 234 - Continued on Next Page

Making Multiple Paced ScansThis program makes multiple scans through a channel list with the scansoccurring at specified intervals. 10 !Dimension a co

Strona 235

Making an Externally Triggered ScanThis example makes one scan through a channel list when the multimeterreceives an external trigger. 10 !Dimension a

Strona 236

Scanning Switchbox Channels (E1326B/E1351A)In this example, the stand-alone multimeter (HP E1326B) scans 5 channelsof an HP E1351A FET multiplexer swi

Strona 237

:VME:STATe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152:VME:STATe? . . . .

Strona 238 - == 0x91)

380 OUTPUT 70903;"INIT"390 !Retrieve the readings from multimeter memory and enter them into the400 !controller. Clear the switchbox to exi

Strona 239

Scanning Switchbox Channels (E1411B/E1460A)In this example, the stand-alone multimeter (HP E1411B) scans 64 channelson an HP E1460A relay multiplexer

Strona 240

370 !Retrieve the readings from multimeter memory, enter and display them380 !on the computer.390 OUTPUT 70903;"FETC?"400 ENTER 70903;Rdgs(*

Strona 241

Multiple High-Speed ScansThis example shows how a scanning multimeter consisting of the HP E1326Bmultimeter and HP E1351A FET multiplexer is programme

Strona 242

370 !Trigger the multimeter to start the measurements. Retrieve the readings380 !from multimeter memory and enter them into the controller. Since the

Strona 243

Maximizing Measurement SpeedThis program shows the multimeter configuration required to makemeasurements at the fastest possible rate (13150 readings/

Strona 244

• The terms MIN and MAX often appear as parameter choices in acommand’s syntax. MIN selects the minimum numeric value for thatparameter. MAX selects

Strona 245

Changing the Data FormatThroughput speed between the multimeter and computer is increased whenbinary (rather than ASCII) data formats are used. The f

Strona 246

• When HP BASIC is used, the program’s ENTER @Dmm USING ...statement is used to remove the Arbitrary Block header:# - tells the computer to terminate

Strona 247

Using a PC, C Language, and theHP 82335 HP-IB Interface CardThe following benchmark program scans 50 channels, 40 times, andcompares each reading to u

Strona 248

:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180:DELay . . . . .

Strona 249

void main(void) /* run the program */ { clrscr(); /* clears screen (turbo C only) */ rst_clr(); /* reset the scanning multimeter */ scan_mu

Strona 250

if (rdgs[j] -.5 || rdgs[j] = .5)printf("\n%f", rdgs[j]);}}}T2 = time(NULL); /* get stop time *//* calculate time for measurements */printf(

Strona 251

Maximizing Measurement AccuracyThis program makes DC voltage measurements on three channels using themultimeter configuration which makes the most acc

Strona 252

Storing Readings in Shared MemoryThe following program stores the multimeter readings on a VME memory card. 10 !Dimension computer variables to store

Strona 253

Comments• Once the INIT command completes, the readings in shared memoryare available to any device. The readings at this time are in 32-bitREAL form

Strona 254

Checking for ErrorsThe following program is a method of checking for errors as you program themultimeter. The program monitors the multimeter’s Stand

Strona 255

Comments• If you have an HP 75000 Series B mainframe with a keyboard, errorscan be monitored by selecting "Monitor" from the multimeter menu

Strona 256

Synchronizing the Multimeter with a ComputerThis is an example of how an HP 9000 Series 200/300 computer canmonitor the multimeter to determine when d

Strona 257

Additional Measurement FunctionsThe following MEASure and CONFigure statements can be substituted intothe example programs to make measurements other

Strona 258

!2-wire thermistor (type = 2252, 5000, 10000).OUTPUT 70903;"MEAS:TEMP? THER,type,(@channel_list)"!4-wire thermistor (type = 2252, 5000, 1000

Strona 259

Reading the Device Type Register . . . . . . . . . . . . . . . . . . . . . . . . . . 224Reading the Query Response Register

Strona 260

Additional FunctionUsing the HP E1345AMultiplexerThis is an example of how to setup scanning when using an HP E1345Amultiplexer configured as a switch

Strona 261 - Useful Tables

310 !320 !Subprogram Scan_100µs.330 !340 SUB Scan_100us350 COM @Sys,@Dvm,@Sw360 DIM Readings(0:15)370 !380 !Clear and reset multimeter.3

Strona 262

850 !Clear and reset multimeter. 860 ! 870 CLEAR @Dvm 880 OUTPUT @Dvm;"*RST;*CLS;*OPC?" 890 ENTER @Dvm;A 900 !

Strona 263 - Programming Error

Notes74 Using the HP E1326B/E1411B Multimeter Chapter 3

Strona 264 - Power-On Settings

Chapter 4Understanding the HP E1326B/E1411BMultimeterAbout This Chapter This chapter describes the parameters which configure the multimeter andhelps

Strona 265 - Aperture Change

Using MEASure and CONFigure CommandsEach time the multimeter makes a measurement, it does so from aconfiguration based on several parameters. The eas

Strona 266 - Interrupt

Table 4-1. Configurations Using MEASure and CONFigureParameter Command SettingFunction As specified.RangeVOLTage:RANGeRESistance:RANGeAs specified or

Strona 267

How to Make MeasurementsThis section explains when you should use MEASure or CONFigure toconfigure the multimeter. It also shows you how to make meas

Strona 268 - Appendix D

Making MeasurementsWhen Using CONFigureTo make a measurement the multimeter must be in the wait-for-trigger statewhen a trigger signal occurs. The ME

Strona 269

Data Formats and DestinationsThe HP E1326B/E1411B multimeter allows you to specify themeasurement (data) format and reading destination parameters whi

Strona 270

CertificationHewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Hewlett-P

Strona 271

Overload Indications The multimeter indicates an overload condition (input greater than theselected range can measure) by displaying or storing:±9.900

Strona 272

Example: EnteringData into the Computer(measurements usingMEASure) 10 !Declare computer array to store 5 readings. 20 REAL Dc_rdgs(1:5) 30 !Configure

Strona 273

4. Data stored in memory overwrites the data from a previous command.5. Each reading stored in memory is four bytes (REAL 32-bit). Thisformat canno

Strona 274 - Decrease Aperture

Example: RetrievingData from Memory 10 !Declare computer array to store 12 readings. 20 REAL Ohm_rdgs(1:12) 30 !Configure the multimeter. 40 OUTPUT 70

Strona 275 - Temperature

ReadingDestinationSummaryThe reading destination you select will depend on your application.However, consider the following when selecting a destinati

Strona 276 - Reserves room for 1000 four

Measurement FunctionsThe HP E1326B/E1411B multimeter can make the following measurements:– DC Voltage– RMS AC Voltage– 2-Wire Resistance– 4-Wire Resis

Strona 277

The AC voltage function is specified as: VOLTage:ACand generally appears in the MEASure and CONFigure commands as:MEAS:VOLT:AC? ... [(@channel_list)]C

Strona 278

Two-WireMeasurementsTwo-wire measurements are useful in applications where test leadresistance is not critical. Because the multimeter measures the t

Strona 279

This function appears in the MEASure and CONFigure commands as:MEAS:TEMP? THER,type,(@channel_list)CONF:TEMP THER,type,(@channel_list)Four-wire measur

Strona 280

Four-wire measurements are specified as:TEMP FRTD,typeThe function appears in the MEASure and CONFigure commands as:MEAS:TEMP? FRTD,type[,(@channel_li

Strona 281

Frame or chassis ground terminal—typi-cally connects to the equipment’s metalframe.Alternating current (AC).Direct current (DC). Indicates hazardous v

Strona 282

Example: ChangingMeasurementFunctions withFUNCtion 10 !Configure for DC voltage measurement. 20 CONF:VOLT:DC 30 !Put multimeter in wait-for-trigger st

Strona 283

Table 4-5. Aperture Time, Range, and Resolution SettingsAperture Time Integration Time (PLCs) 10 µs*0.0005100 µs0.0052.5 ms0.12516.7 ms120 ms1267 ms16

Strona 284

Range The range parameter sets the range of input signal levels the multimeter isto accept and measure. Consider the following when determining a ran

Strona 285

Autorange The default range is autorange. Autorange is the process where themultimeter samples the input signal, and then automatically selects theco

Strona 286

Querying theAutorange SettingThe autorange setting is queried with the VOLTage:RANGe:AUTO? andRESistance:RANGe:AUTO? commands. See Chapter 5 for addi

Strona 287

Setting the Resolution The resolutions for DC/AC voltage and resistance measurements are givenin Table 4-5 on page 92. Note that the resolution is sp

Strona 288

Aperture andIntegration TimeThe aperture time or integration time is the time which the multimetersamples the input signal. Aperture time is expresse

Strona 289

Setting the Apertureand Integration TimeThe multimeter aperture times, integration times, line frequency rejected,and the amount of normal mode reject

Strona 290

Autozero Autozero is the process of cancelling out the offset voltage from DC voltageand resistance measurements. When the multimeter is triggered an

Strona 291

OffsetCompensationAnytime a resistance measurement is made, offset voltages internal andexternal to the multimeter can be present. When these offsets