2- 10 Agilent B2900 Users Guide, Edition 2
Introduction
Source/Measure Unit
Measurement Parameters
B2900 supports the following measurement parameters.
Current
Voltage
Resistance
•Power
Resistance data is given by Resistance=Vmeas/Imeas.
Power data is given by Power=VmeasImeas.
In the above formula, Vmeas is the voltage measurement data, and Imeas is the
current measurement data.
For using the resistance compensation, see “Resistance Compensation” on page
6-16.
Limit/Compliance
B2900 has a compliance feature that limits the output voltage or current to prevent
damage to the device under test. When the SMU applies voltage, you can specify the
current limit/compliance. When the SMU applies current, you can specify the
voltage limit/compliance. For details about limit/compliance, see
“Limit/Compliance” on page 6-3.