Agilent Technologies Agilent B2201A Instrukcja Użytkownika Strona 150

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 234
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 149
5-24 Agilent B2200 Users Guide, Edition 2
SCPI Command Reference
:DIAGnostic:TEST:FRAMe:STATe?
:DIAGnostic:TEST:FRAMe:STATe?
This command returns the most recent test result (pass/fail) of the specified test.
Syntax :DIAGnostic:TEST:FRAMe:STATe? CONTroller | FPANel | LED
| PEN | BEEPer
Query response test_result <newline><^END>
1: fail
0: pass
-1: not tested
Example OUTPUT @Agb2200;":DIAG:TEST:FRAM:STAT? CONT"
ENTER @Agb2200;A
[:ROUTe]:AGND:CHANnel:DISable:CARD
This command ground-disables the all output ports (channels) for the specified card.
When the Ground Mode is ON, the ground-disabled output ports are disconnected
from the input Ground Port. At *RST, no channel is ground-enabled. The ground
mode is set by “[:ROUTe]:AGND[:STATe]”.
Syntax [:ROUTe]:AGND:CHANnel:DISable:CARD card_number
Example OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS:CARD ALL"
Parameter Explanation
CONTroller Controller test
FPANel Front panel interface test
LED LED matrix test
PEN Light pen test
BEEPer Beeper test
Parameter Explanation
card_number For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
Przeglądanie stron 149
1 2 ... 145 146 147 148 149 150 151 152 153 154 155 ... 233 234

Komentarze do niniejszej Instrukcji

Brak uwag