Agilent Technologies Agilent B2201A Instrukcja Użytkownika Strona 135

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 234
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 134
Agilent B2200 Users Guide, Edition 2 5-9
SCPI Command Reference
*RST
*RST
This command performs an instrument reset. Status after *RST is shown below:
Channel Configuration: Auto configuration mode
Connection Rule: Free
Connection Sequence: Break Before Make
Bias Mode: Off
Bias Input Port: 10
Bias-enabled Channels: All channels (output ports) are bias-enabled.
Couple Mode: Off
Couple Input Port: Cleared.
Ground Mode: Off
Ground Input Port: 12
Ground-enabled Channels: Cleared.
Ground-enabled Ports: Cleared.
Card Channel Status: All relays are opened.
Self-Test Result: Not changed.
Input Port Symbol String: Cleared.
Output Port Symbol String: Cleared.
Beeper: On
LCD in remote mode: Off
LED in remote mode: On
Light Pen: Enabled.
Syntax *RST
Example OUTPUT @Agb2200;"*RST"
NOTE This command does not change the self-test result.
For the power-on settings, see :SYSTem:CPON.
Przeglądanie stron 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 233 234

Komentarze do niniejszej Instrukcji

Brak uwag