Impedance Measurement (Option 4396B-010)
Measurement functions
Measurement parameters . . . . . . . . . . . . . . . . . . . . . . .Z, Y, L, C, Q, R, X, G, B, θ
Display parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . .[Z], θ
z
, R, X, [Y], θ
y
, G, B,
[Γ], θ
g
, Γ
x
, Γ
y
, Cp, Cs, Lp, Ls, Rp, Rs, D, Q
Display formats
Vertical lin/log scale
Complex plane
Polar/Smith/admittance chart
Sweep parameters
Linear frequency sweep
Logarithmic frequency sweep
List frequency sweep
Linear power sweep (dBm)
IF bandwidth
10, 30, 100, 300,1 k, 3 k, 10 k, 40 k [Hz]
Calibration
OPEN/SHORT/LOAD 3 term calibration
Fixture compensation
Port extension correction
Unknown port
7-mm connector
Output characteristics
1
Frequency range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 kHz to 1.8 GHz
Frequency resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 mHz
Output level . . . . . . . . . . . . . . . . . . . . . . . . –60 to +20 dBm (@RF OUT port)
Output level accuracy . . . . . . . . . . . . . . . . . . . A + B + 6 [dB] x F/(1.8 x 10
9
)
Where,
A = 2 dB (±5 °C)
B = 0 dB(GSC 0 dBm ), or 1 dB (–40 GSC < 0 dBm ),
or 2 dB (–60 GSC < –40 dBm )
F is output frequency.
Output level resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1 dB
Measurement port impedance . . . . . . . . . . . . . . . . . . . . . . . . . .Nominal 50
1. Signal level at the measurement port is 6 dB lower than the RF GUT port
when the measurement port is terminated by 50 .