4- 26 Agilent B2900 Users Guide, Edition 2
Front Panel Reference
Config key group
Output Filter dialog box
This dialog box provides the following parameters for setting the output filter.
Ch Only on 2-channel models. Channel 1 (Ch 1) or 2 (Ch 2)
This field specifies the channel set by this dialog box.
Filter State Output filter, ON or OFF
Set the filter ON to obtain clean source output without spikes
and overshooting. Note, however, that using a filter may
increase the SMU settling time.
Automatic Filter Automatic filter, ON or OFF
Set the function ON to automatically set the output filter which
provides the best filter characteristics and cutoff frequency.
Time Constant Filter time constant, 5 s to 500 s