Installation Note 83650-90007 15
Reassembly
54. Replace the front panel in the frame.
55. Replace the two screws on the lower front frame edge.
56. Replace the three screws on the top front frame edge.
57. Replace the instrument side covers and top and bottom covers. (Be sure to replace the RF braid in
the side channels).
58. Replace the four back feet and four bottom feet.
59. Replace the vinyl trim strip on the front edge.
Hardware Configuration Number
60. The calibration constant for the hardware configuration must be updated to reflect the new
configuration. Change the calibration constant and store the constants to memory. If you are
unfamiliar with storing calibration constants, instructions are provided in the “Calibration
Constants” chapter of the Service Guide.
Cal Constant: #463
Value: +4096 (adding Option 004)
Operation Verification
61. Perform the following adjustments and performance tests.
NOTE If any failures occur, perform the Step Attenuator Flatness adjustment (auto) and the Step
Attenuator Flatness Test (auto).
This completes the installation.
Adjustments Performance Tests
None Full Self-Test
Power Flatness